{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79682",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution. US nationwide, Netherlands, Mexico, Paraguay, Brazil, Argentina, Malaysia, Costa Rica, Chile, South Korea, and Canada.",
      "recall_number": "Z-1734-2018",
      "product_description": "Gentle Threads, Interference Screw; Model No. 905617, for soft tissue reattachment procedures in the ankle/foot and knee.",
      "product_quantity": "54",
      "reason_for_recall": "Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.",
      "recall_initiation_date": "20180328",
      "center_classification_date": "20180508",
      "termination_date": "20190329",
      "report_date": "20180516",
      "code_info": "lot 207910"
    }
  ]
}