  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68271",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution including US Nationwide, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.",
      "recall_number": "Z-1734-2014",
      "product_description": "VITROS  4600 Chemistry System, Product Code/Catalog Number 6802445   for in vitro diagnostic use.",
      "product_quantity": "Domestic - 82, Foreign - 213 units",
      "reason_for_recall": "Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS  4600 Chemistry System, the VITROS 5600  Integrated System and the VITROS  5,1 Chemistry System. MW codes may be associated with results from any of the following VITROS Chemistry Products using multiple point (multipoint) rate measurements. The MicroSlide assays that could be affected",
      "recall_initiation_date": "20140410",
      "center_classification_date": "20140609",
      "termination_date": "20180521",
      "report_date": "20140618",
      "code_info": "Software Version 3.0 and lower; Serial Numbers J46000108 through J46000403"
    }
  ]
}