  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "A small number of Vista Air Compressors were miswired, which if installed and powered up, could cause a fire, or become an electrical shock hazard.",
      "address_2": "PO BOX 6101",
      "product_quantity": "4",
      "code_info": "Product Siemens Material Number (SMN)  Dimension Vista 500 Domestic: 10284473  International: 10488224  Dimension Vista 1500 Domestic: 10444801  International: 10444802  Serial #\u0019s: 0913900306 \u0013 0913900355",
      "center_classification_date": "20140609",
      "distribution_pattern": "Worldwide Distribution: US Distribution to states of: IL, OH and TN; and country of: Italy.",
      "state": "DE",
      "product_description": "Siemens Healthcare Diagnostics Vista Air Compressor.     Vista AirCompressor replacement part: SMN 10706557  SKIT Air Compressor Assy, used on the Dimension Vista 500 and Dimension Vista 1500 systems.    Siemens Healthcare Diagnostics Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric and integrated ion-selective multisensory technology for clinical use.",
      "report_date": "20140618",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-1733-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68076",
      "termination_date": "20141215",
      "recall_initiation_date": "20140414",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}