  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "60480",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines,  Singapore, Taiwan, and Thailand.",
      "recall_number": "Z-1732-2013",
      "product_description": "CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules.  Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.",
      "product_quantity": "3972 since 4/26/12",
      "reason_for_recall": "Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.",
      "recall_initiation_date": "20130227",
      "center_classification_date": "20130713",
      "termination_date": "20160217",
      "report_date": "20130724",
      "code_info": "Catalog number: 500AV* and serial numbers: 0024, 0033, 0040, 0051, 1192, 1193, 1355,1356, 1420, 1423, 1464, 1465, 1467, 1468,1483,1504,1521-1524,1526,1527,1529,1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901,2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, 5578, and 5579."
    }
  ]
}