{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70918",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide, Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.",
      "recall_number": "Z-1729-2015",
      "product_description": "Assay Data Disk (ADD), All DRV versions that support product within expiry dating    Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use. VITROS Urine Assays: VITROS Calcium (Ca), Magnesium (Mg), and Phosphorus (Phos) requires acidification pretreatment prior to urine assay testing",
      "product_quantity": "2,387 units Total   (987 domestically & 1400 internationally)",
      "reason_for_recall": "Software anomaly regarding urine samples that require acidified pretreatment.  The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.",
      "recall_initiation_date": "20150406",
      "center_classification_date": "20150608",
      "termination_date": "20180525",
      "report_date": "20150617",
      "code_info": "Product Code: 6801876  Unique Device Identifier Numbers: 10758750001576"
    }
  ]
}