{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92328",
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution",
      "recall_number": "Z-1725-2023",
      "product_description": "ICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1;  Implantable Cardioverter Defibrillators",
      "product_quantity": "1941 units",
      "reason_for_recall": "There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.",
      "recall_initiation_date": "20230510",
      "center_classification_date": "20230616",
      "report_date": "20230628",
      "code_info": "GTIN 00763000178376, Lot Serial Numbers:  RSN600341S, RSN601060S, RSN601316S, RSN601817S, RSN602193S, RSN600668S, RSN600669S, RSN600670S, RSN600672S, RSN601804S, RSN601808S, RSN601810S, RSN602514S, RSN601050S, RSN601241S, RSN602551S, RSN600820S, RSN603061S, RSN601870S, RSN602495S, RSN601384S, RSN601691S, RSN601237S, RSN602951S, RSN603065S, RSN600626S, RSN600877S, RSN601019S, RSN601020S, RSN601022S, RSN601027S, RSN601028S, RSN601029S, RSN601030S, RSN601032S, RSN601038S, RSN601040S, RSN600997S, RSN601013S, RSN601039S, RSN600837S, RSN601913S, RSN601957S, RSN600334S, RSN600116S, RSN602781S, RSN602004S, RSN602073S, RSN601983S, RSN600628S, RSN601374S, RSN600624S, RSN601004S, RSN601902S, RSN602210S, RSN600402S, RSN600359S, RSN600360S, RSN600361S, RSN600366S, RSN601283S, RSN602071S, RSN602182S, RSN602365S, RSN602756S, RSN602171S, RSN600482S, RSN600870S, RSN601950S, RSN602490S, RSN600952S, RSN602026S, RSN602072S, RSN602244S, RSN602297S, RSN600640S, RSN601351S, RSN601352S, RSN601830S, RSN601831S, RSN600813S, RSN600891S, RSN601202S, RSN601390S, RSN602166S, RSN602167S, RSN602245S, RSN600550S, RSN601897S, RSN602283S, RSN602325S, RSN602485S, RSN600459S, RSN600461S, RSN600463S, RSN600574S, RSN600639S, RSN600918S, RSN601080S, RSN601116S, RSN601140S, RSN601152S, RSN601207S, RSN601302S, RSN601694S, RSN602029S, RSN602194S, RSN600673S, RSN601354S, RSN602346S, RSN601149S, RSN601829S, RSN601832S, RSN601874S, RSN600522S, RSN602059S, RSN602379S, RSN602424S, RSN601230S, RSN601402S, RSN602202S, RSN602397S, RSN602539S, RSN602506S, RSN602342S, RSN602901S, RSN602221S, RSN601269S, RSN600766S, RSN600803S, RSN600091S, RSN600345S, RSN601932S, RSN600736S, RSN600775S, RSN600846S, RSN600949S, RSN601186S, RSN601191S, RSN601270S, RSN601476S, RSN602049S, RSN600098S, RSN600452S, RSN600646S, RSN602201S, RSN602750S, RSN601279S, RSN602343S, RSN602947S, RSN601991S, RSN601199S, RSN602199S, RSN600441S, RSN600682S, RSN602316S, RSN602688S, RSN602693S, RSN600833S, RSN601882S, RSN602767S, RSN601330S, RSN600854S, RSN601242S, RSN601266S, RSN601437S, RSN602063S, RSN600566S, RSN602719S, RSN600418S, RSN600660S, RSN600968S, RSN601190S, RSN601365S, RSN602887S, RSN602964S, RSN602685S, RSN600714S, RSN602449S, RSN602529S, RSN600223S, RSN602062S, RSN602150S, RSN600778S, RSN600780S, RSN601974S, RSN602255S, RSN600440S, RSN601907S, RSN601327S, RSN601399S, RSN600627S, RSN600860S, RSN600871S, RSN600937S, RSN601177S, RSN601837S, RSN601924S, RSN602009S, RSN602348S, RSN602360S, RSN601382S, RSN602306S, RSN602321S, RSN600329S, RSN600843S, RSN600879S, RSN602771S, RSN600523S, RSN600537S, RSN601155S, RSN601169S, RSN602370S, RSN602434S, RSN601383S, RSN602310S, RSN602178S, RSN600994S, RSN601182S, RSN601754S, RSN600145S, RSN600569S, RSN600925S, RSN600926S, RSN601171S, RSN601193S, RSN601400S, RSN602056S, RSN602192S, RSN600916S, RSN601223S, RSN601705S, RSN601930S, RSN602285S, RSN600218S, RSN601871S, RSN602482S, RSN601333S, RSN601716S, RSN601719S, RSN601802S, RSN602702S, RSN603066S, RSN602274S, RSN602694S, RSN600688S, RSN601044S, RSN601730S, RSN601748S, RSN601750S, RSN601753S, RSN602227S, RSN602544S, RSN601204S, RSN601951S, RSN601309S, RSN601833S, RSN601834S, RSN600225S, RSN600269S, RSN601350S, RSN600587S, RSN600589S, RSN600594S, RSN600643S, RSN600796S, RSN600939S, RSN601008S, RSN601943S, RSN601996S, RSN602252S, RSN602384S, RSN602229S, RSN601319S, RSN600542S, RSN600543S, RSN600772S, RSN601315S, RSN602294S, RSN600095S, RSN600535S, RSN600912S, RSN601271S, RSN602005S, RSN602139S, RSN602335S, RSN602387S, RSN602760S, RSN600807S, RSN601011S, RSN601334S, RSN602195S, RSN602483S, RSN602511S, RSN600212S, RSN600454S, RSN600653S, RSN600794S, RSN600806S, RSN602660S, RSN602883S, RSN602462S, RSN602463S, RSN602464S, RSN602804S, RSN602876S, RSN602877S, RSN602878S, RSN602882S, RSN602884S, RSN602885S, RSN600431S, RSN600598S, RSN601212S, RSN601308S, RSN601362S, RSN601377S, RSN602410S, RSN600413S, RSN600827S, RSN600882S, RSN601299S, RSN602319S, RSN600802S, RSN602765S, RSN602769S, RSN602324S, RSN602340S, RSN602930S, RSN602962S, RSN600215S, RSN600216S, RSN600744S, RSN600757S, RSN600972S, RSN601068S, RSN602133S, RSN602516S, RSN600808S, RSN600291S, RSN600448S, RSN600761S, RSN602448S, RSN600384S, RSN601265S, RSN602938S, RSN602531S, RSN600122S, RSN602749S, RSN600505S, RSN600657S, RSN600662S, RSN600770S, RSN601906S, RSN600353S, RSN600419S, RSN601076S, RSN601146S, RSN600421S, RSN602020S, RSN600765S, RSN601431S, RSN600379S, RSN600407S, RSN600661S, RSN602180S, RSN602393S, RSN600134S, RSN601272S, RSN602409S, RSN601721S, RSN601479S, RSN600671S, RSN601921S, RSN602117S, RSN602429S, RSN600387S, RSN602593S, RSN600150S, RSN600540S, RSN601066S, RSN601908S, RSN601939S, RSN602292S, RSN602008S, RSN602720S, RSN600370S, RSN600773S, RSN601141S, RSN601903S, RSN602414S, RSN602466S, RSN600804S, RSN600805S, RSN600978S, RSN601058S, RSN601084S, RSN601398S, RSN601961S, RSN602533S, RSN600781S, RSN601329S, RSN602030S, RSN602419S, RSN602663S, RSN600130S, RSN600169S, RSN600266S, RSN600372S, RSN600411S, RSN600503S, RSN600558S, RSN600586S, RSN600911S, RSN600923S, RSN601107S, RSN601463S, RSN601910S, RSN601954S, RSN602190S, RSN602351S, RSN602499S, RSN600151S, RSN600991S, RSN602899S, RSN602900S, RSN600619S, RSN600749S, RSN600750S, RSN600788S, RSN600822S, RSN600830S, RSN601391S, RSN601425S, RSN601471S, RSN602077S, RSN602512S, RSN600330S, RSN600416S, RSN600462S, RSN600985S, RSN601091S, RSN601094S, RSN601131S, RSN602173S, RSN602238S, RSN601771S, RSN602213S, RSN602699S, RSN602897S, RSN600144S, RSN602215S, RSN602540S, RSN600209S, RSN600340S, RSN600464S, RSN601372S, RSN601918S, RSN600651S, RSN601419S, RSN602328S, RSN600616S, RSN601073S, RSN601695S, RSN600623S, RSN601119S, RSN601209S, RSN602358S, RSN602378S, RSN602494S, RSN600162S, RSN601875S, RSN602339S, RSN602643S, RSN600131S, RSN602320S, RSN602584S, RSN601249S, RSN601847S, RSN602039S, RSN600365S, RSN600631S, RSN600715S, RSN601461S, RSN600458S, RSN602257S, RSN602687S, RSN601328S, RSN600331S, RSN600609S, RSN600969S, RSN602677S, RSN602678S, RSN602785S, RSN601949S, RSN601967S, RSN601972S, RSN602031S, RSN602032S, RSN602033S, RSN602556S, RSN602559S, RSN602574S, RSN602902S, RSN602903S, RSN602906S, RSN601007S, RSN601766S, RSN601885S, RSN602137S, RSN602787S, RSN600017S, RSN602145S, RSN600734S, RSN600740S, RSN600763S, RSN600826S, RSN600853S, RSN601105S, RSN601395S, RSN602119S, RSN602504S, RSN602870S, RSN602152S, RSN602653S, RSN600148S, RSN600466S, RSN600514S, RSN600851S, RSN601236S, RSN601966S, RSN602120S, RSN602550S, RSN600446S, RSN600934S, RSN601336S, RSN600219S, RSN600641S, RSN602042S, RSN600202S, RSN601931S, RSN600149S, RSN600595S, RSN602299S, RSN601208S, RSN602012S, RSN602172S, RSN602301S, RSN600637S, RSN600333S, RSN601878S, RSN601889S, RSN601923S, RSN602394S, RSN600979S, RSN601111S, RSN602477S, RSN602262S, RSN602022S, RSN602023S, RSN602024S, RSN602696S, RSN600638S, RSN602331S, RSN602698S, RSN600435S, RSN600521S, RSN600548S, RSN600621S, RSN600703S, RSN600762S, RSN600818S, RSN601082S, RSN601394S, RSN601962S, RSN602143S, RSN602162S, RSN602212S, RSN602350S, RSN602421S, RSN602553S, RSN602558S, RSN601816S, RSN602398S, RSN601153S, RSN601368S, RSN602513S, RSN602125S, RSN602745S, RSN600848S, RSN601064S, RSN601065S, RSN601165S, RSN601393S, RSN601407S, RSN602149S, RSN602183S, RSN602714S, RSN602917S, RSN600490S, RSN600568S, RSN600675S, RSN600981S, RSN600984S, RSN601074S, RSN601075S, RSN601427S, RSN601439S, RSN602027S, RSN602134S, RSN602165S, RSN602241S, RSN602359S, RSN602446S, RSN602447S, RSN602488S, RSN602502S, RSN602151S, RSN602420S, RSN600530S, RSN601063S, RSN601881S, RSN602444S, RSN602918S, RSN602730S, RSN600197S, RSN600389S, RSN600559S, RSN601168S, RSN602552S, RSN602892S, RSN602908S, RSN601081S, RSN601342S, RSN602204S, RSN602909S, RSN600394S, RSN600555S, RSN600894S, RSN600951S, RSN600995S, RSN601100S, RSN601121S, RSN601124S, RSN601396S, RSN601403S, RSN601434S, RSN602138S, RSN602144S, RSN602249S, RSN602314S, RSN602352S, RSN600203S, RSN600845S, RSN602761S, RSN600371S, RSN601228S, RSN600527S, RSN601000S, RSN602729S, RSN602753S, RSN602015S, RSN600207S, RSN601345S, RSN601976S, RSN602910S, RSN602911S, RSN602913S, RSN602914S, RSN600439S, RSN600745S, RSN602135S, RSN602337S, RSN602959S, RSN600999S, RSN602146S, RSN602341S, RSN602479S, RSN600509S, RSN600533S, RSN601981S, RSN602086S, RSN601232S, RSN601998S, RSN602132S, RSN602778S, RSN600942S, RSN601185S, RSN601373S, RSN601843S, RSN602174S, RSN601287S, RSN600488S, RSN602214S, RSN601340S, RSN602704S, RSN602366S, RSN601929S, RSN602066S, RSN602069S, RSN602114S, RSN602571S, RSN602705S, RSN600743S, RSN600348S, RSN601159S, RSN602128S, RSN600874S, RSN600954S, RSN600955S, RSN601246S, RSN601301S, RSN601335S, RSN601356S, RSN601417S, RSN601876S, RSN602519S, RSN602520S, RSN601136S, RSN602807S, RSN602810S, RSN602872S, RSN602873S, RSN602874S, RSN602864S, RSN602871S, RSN602973S, RSN602707S, RSN600401S, RSN600685S, RSN600768S, RSN600980S, RSN601324S, RSN601404S, RSN601466S, RSN601879S, RSN601925S, RSN601948S, RSN602060S, RSN602188S, RSN602276S, RSN600831S, RSN600382S, RSN600390S, RSN602895S, RSN602723S, RSN602391S, RSN602541S, RSN600810S, RSN601042S, RSN601322S, RSN602356S, RSN601135S, RSN601260S, RSN602040S, RSN602442S, RSN600504S, RSN600676S, RSN600104S, RSN600647S, RSN600363S, RSN600719S, RSN600847S, RSN601142S, RSN602115S, RSN601263S, RSN600112S, RSN600161S, RSN601056S, RSN601346S, RSN601347S, RSN601378S, RSN601442S, RSN601472S, RSN601898S, RSN602028S, RSN601198S, RSN600473S, RSN601125S, RSN601126S, RSN601129S, RSN601253S, RSN601254S, RSN601255S, RSN601421S, RSN601426S, RSN601428S, RSN600575S, RSN601343S, RSN601864S, RSN601978S, RSN602354S, RSN602425S, RSN602428S, RSN600689S, RSN600725S, RSN600727S, RSN601409S, RSN601732S, RSN601883S, RSN602158S, RSN602263S, RSN601975S, RSN602390S, RSN602726S, RSN602727S, RSN602728S, RSN600930S, RSN602555S, RSN600487S, RSN600590S, RSN600591S, RSN601430S, RSN602822S, RSN600604S, RSN602246S, RSN601118S, RSN600205S, RSN602537S, RSN600111S, RSN600113S, RSN600352S, RSN600362S, RSN602748S, RSN600378S, RSN601703S, RSN601741S, RSN602416S, RSN600835S, RSN601762S, RSN601977S, RSN602268S, RSN602269S, RSN602270S, RSN601867S, RSN601239S, RSN602780S, RSN600914S, RSN602065S, RSN602247S, RSN602329S, RSN600332S, RSN600350S, RSN600611S, RSN600888S, RSN600953S, RSN601755S, RSN600194S, RSN600759S, RSN601797S, RSN602043S, RSN600889S, RSN600921S, RSN601405S, RSN601009S, RSN602303S, RSN602440S, RSN600453S, RSN600573S, RSN602435S, RSN600986S, RSN602960S, RSN600126S, RSN602543S, RSN602695S, RSN600127S, RSN600414S, RSN601031S, RSN602279S, RSN602290S, RSN602326S, RSN602493S, RSN602939S, RSN600128S, RSN600706S, RSN600718S, RSN600947S, RSN600975S, RSN601148S, RSN601276S, RSN601401S, RSN601922S, RSN602052S, RSN602057S, RSN600189S, RSN600436S, RSN600881S, RSN601240S, RSN601786S, RSN601964S, RSN600525S, RSN601304S, RSN601067S, RSN601749S, RSN600460S, RSN600648S, RSN600700S, RSN600739S, RSN600752S, RSN600815S, RSN600885S, RSN601325S, RSN600760S, RSN601109S, RSN602651S, RSN602690S, RSN600605S, RSN601273S, RSN602363S, RSN602528S, RSN602226S, RSN602265S, RSN602441S, RSN602503S, RSN602564S, RSN603063S, RSN600841S, RSN600895S, RSN600904S, RSN600905S, RSN600906S, RSN600907S, RSN600909S, RSN602260S, RSN602967S, RSN600588S, RSN601332S, RSN602267S, RSN602548S, RSN602570S, RSN602572S, RSN600582S, RSN600570S, RSN601130S, RSN600395S, RSN600409S, RSN602580S, RSN602547S, RSN600221S, RSN602136S, RSN602703S, RSN601724S, RSN602231S, RSN602353S, RSN600825S, RSN603067S, RSN602453S, RSN600892S, RSN600351S, RSN601480S, RSN602725S, RSN601901S, RSN602000S, RSN600153S, RSN600883S, RSN600897S, RSN601137S, RSN601238S, RSN601936S, RSN602588S, RSN600173S, RSN600417S, RSN601133S, RSN601245S, RSN602470S, RSN600608S, RSN600613S, RSN600684S, RSN601092S, RSN601303S, RSN602396S, RSN601965S, RSN602716S, RSN601955S, RSN601956S, RSN601994S, RSN601995S, RSN601997S, RSN602076S, RSN602527S, RSN601890S, RSN601893S, RSN600976S, RSN602258S, RSN602432S, RSN600328S, RSN600901S, RSN600424S, RSN600741S, RSN600856S, RSN600862S, RSN600872S, RSN601353S, RSN601429S, RSN602035S, RSN602129S, RSN602304S, RSN602389S, RSN600180S, RSN600615S, RSN600893S, RSN601183S, RSN601057S, RSN602438S, RSN602935S, RSN602203S, RSN602223S, RSN602308S, RSN602381S, RSN600349S, RSN601158S, RSN602191S, RSN600839S, RSN602706S, RSN602472S, RSN600597S, RSN600484S, RSN601275S, RSN601306S, RSN600208S, RSN602649S, RSN602691S, RSN603058S, RSN600927S, RSN600929S, RSN600931S, RSN601267S, RSN602240S, RSN602284S, RSN602362S, RSN602443S, RSN602524S, RSN602532S, RSN602560S, RSN600233S, RSN602013S, RSN602045S, RSN602047S, RSN602048S, RSN601095S, RSN602675S, RSN602654S, RSN602655S, RSN602656S, RSN602657S, RSN602676S, RSN600922S, RSN602415S, RSN600373S, RSN600374S, RSN602496S, RSN602946S, RSN600698S, RSN600920S, RSN602368S, RSN600919S, RSN601106S, RSN601363S, RSN602148S, RSN600191S, RSN600406S, RSN601285S, RSN601360S, RSN600832S, RSN602287S, RSN600563S, RSN601312S, RSN600472S, RSN601172S, RSN601173S, RSN601927S, RSN602273S, RSN600720S, RSN602006S, RSN600699S, RSN602715S, RSN600483S, RSN600499S, RSN600932S, RSN601023S, RSN601087S, RSN601375S, RSN601422S, RSN602037S, RSN602087S, RSN602206S, RSN602431S, RSN602436S, RSN600680S, RSN602034S, RSN602312S, RSN602437S, RSN602475S, RSN602217S, RSN602242S, RSN602334S, RSN602963S, RSN602865S, RSN600430S, RSN600437S, RSN601947S, RSN601221S, RSN602272S, RSN602430S, RSN600553S, RSN602896S, RSN602710S, RSN602445S, RSN601973S, RSN600679S, RSN600797S, RSN601132S, RSN601205S, RSN601321S, RSN602777S, RSN600174S, RSN600502S, RSN600520S, RSN600528S, RSN600541S, RSN600600S, RSN600814S, RSN601096S, RSN601163S, RSN601196S, RSN601200S, RSN601307S, RSN601704S, RSN601886S, RSN601944S, RSN602373S, RSN602880S, RSN602881S, RSN600376S, RSN600683S, RSN601905S, RSN602154S, RSN602256S, RSN600771S, RSN601767S, RSN602330S, RSN602478S, RSN600834S, RSN601725S, RSN601892S, RSN601928S, RSN601935S, RSN601999S, RSN602747S, RSN602746S, RSN600983S, RSN602549S, RSN600677S, RSN601256S, RSN601474S, RSN602517S, RSN600510S, RSN601381S, RSN600811S, RSN601123S, RSN600545S, RSN602793S, RSN602796S, RSN600099S, RSN602647S, RSN602709S, RSN600388S, RSN600444S, RSN600524S, RSN600823S, RSN600865S, RSN601261S, RSN601288S, RSN601339S, RSN601358S, RSN601470S, RSN602046S, RSN602116S, RSN602121S, RSN602130S, RSN602349S, RSN602469S, RSN602492S, RSN602509S, RSN602568S, RSN602585S, RSN602586S, RSN602594S, RSN602869S, RSN602866S, RSN600910S, RSN600782S, RSN600650S, RSN603076S, RSN600129S, RSN602577S, RSN600434S, RSN601318S, RSN601846S, RSN602575S, RSN600844S, RSN602372S, RSN602386S, RSN602423S, RSN601061S, RSN602147S, RSN600135S, RSN600171S, RSN600179S, RSN600220S, RSN600567S, RSN600572S, RSN600654S, RSN600748S, RSN600840S, RSN600880S, RSN600990S, RSN600993S, RSN601002S, RSN601128S, RSN601313S, RSN601478S, RSN601946S, RSN602278S, RSN602402S, RSN602567S, RSN602569S, RSN600163S, RSN600175S, RSN600456S, RSN600663S, RSN601166S, RSN602197S, RSN600571S, RSN600583S, RSN600859S, RSN600666S, RSN602067S, RSN602361S, RSN600764S, RSN600829S, RSN602205S, RSN602336S, RSN602701S, RSN600399S, RSN602307S, RSN600896S, RSN601317S, RSN602187S, RSN602929S, RSN602927S, RSN602254S, RSN600375S, RSN601194S, RSN600938S, RSN601117S, RSN600581S, RSN602207S, RSN602412S, RSN601093S, RSN600117S, RSN602374S, RSN602686S, RSN600090S, RSN600198S, RSN600199S, RSN600204S, RSN600347S, RSN600380S, RSN600433S, RSN600476S, RSN600489S, RSN600868S, RSN600964S, RSN601018S, RSN601045S, RSN601077S, RSN601224S, RSN601934S, RSN601979S, RSN602250S, RSN602298S, RSN602399S, RSN600495S, RSN600511S, RSN601156S, RSN600792S, RSN602118S, RSN602159S, RSN602664S, RSN602722S, RSN602007S, RSN600124S, RSN601899S, RSN600474S, RSN600747S, RSN600774S, RSN600776S, RSN602186S, RSN602019S, RSN602454S, RSN600928S, RSN600343S, RSN600584S, RSN601234S, RSN600121S, RSN601181S, RSN602680S, RSN602681S, RSN600120S, RSN600167S, RSN600438S, RSN602021S, RSN602050S, RSN600139S, RSN600391S, RSN600966S, RSN602175S, RSN602251S, RSN601278S, RSN602666S, RSN602208S, RSN602811S, RSN602545S, RSN602589S, RSN602724S, RSN601475S, RSN602142S, RSN602798S, RSN600915S, RSN601035S, RSN601046S, RSN601048S, RSN601049S, RSN602157S, RSN602161S, RSN602264S, RSN602296S, RSN602426S, RSN602450S, RSN602562S, RSN602782S, RSN602783S, RSN602784S, RSN602800S, RSN602801S, RSN602802S, RSN602888S, RSN600108S, RSN600109S, RSN600123S, RSN600987S, RSN602452S, RSN601880S, RSN602332S, RSN601958S, RSN601015S, RSN601016S, RSN601108S, RSN601115S, RSN601693S, RSN602774S, RSN600486S, RSN601835S, RSN602405S, RSN602413S, RSN602155S, RSN602669S, RSN600423S, RSN601971S, RSN602526S, RSN602789S, RSN602791S, RSN600196S, RSN602788S, RSN600547S, RSN602646S, RSN601034S, RSN601244S, RSN601348S, RSN601839S, RSN601841S, RSN602179S, RSN602309S, RSN602382S, RSN602484S, RSN601114S, RSN602581S, RSN601059S, RSN601139S, RSN602538S, RSN601055S, RSN601014S, RSN601264S, RSN600107S, RSN601071S, RSN601179S, RSN602184S, RSN600183S, RSN600549S, RSN601097S, RSN600342S, RSN600356S, RSN600422S, RSN600447S, RSN600465S, RSN601435S, RSN601441S, RSN600200S, RSN600593S, RSN600665S, RSN600819S, RSN601078S, RSN601251S, RSN600753S, RSN601379S, RSN601387S, RSN601157S, RSN601143S, RSN601827S, RSN602036S, RSN602300S, RSN600965S, RSN602271S, RSN601062S, RSN602282S, RSN601380S, RSN602717S, RSN602718S, RSN602889S, RSN602894S, RSN603078S, RSN600185S, RSN600192S, RSN600195S, RSN600201S, RSN600344S, RSN600358S, RSN600429S, RSN600544S, RSN600561S, RSN600690S, RSN600754S, RSN600977S, RSN601025S, RSN601098S, RSN601367S, RSN601376S, RSN601900S, RSN602044S, RSN602068S, RSN602160S, RSN602461S, RSN602508S, RSN602530S, RSN600789S, RSN600982S, RSN601053S, RSN601215S, RSN602465S, RSN600884S, RSN602323S, RSN602713S, RSN601033S, RSN601941S, RSN601916S, RSN601216S, RSN600338S, RSN600728S, RSN600793S, RSN600946S, RSN601277S, RSN600443S, RSN601990S, RSN600516S, RSN602697S, RSN601113S, RSN602375S, RSN600817S, RSN600539S, RSN600876S, RSN602683S, RSN600645S, RSN601174S, RSN601247S, RSN601369S, RSN601772S, RSN602456S, RSN601085S, RSN601201S, RSN601408S, RSN601411S, RSN601438S, RSN602369S, RSN600157S, RSN600186S, RSN600451S, RSN600578S, RSN601134S, RSN601706S, RSN601915S, RSN601940S, RSN601970S, RSN600580S, RSN602189S, RSN600410S, RSN600368S, RSN600492S, RSN601873S, RSN602011S, RSN602295S, RSN600146S, RSN600147S, RSN601175S, RSN600427S, RSN600428S, RSN600506S, RSN600515S, RSN600642S, RSN600957S, RSN602395S, RSN600467S, RSN600468S, RSN600469S, RSN600470S, RSN600471S, RSN600477S, RSN600479S, RSN600517S, RSN600842S, RSN601072S, RSN601229S, RSN601258S, RSN601268S, RSN601311S, RSN601320S, RSN601709S, RSN601896S, RSN601911S, RSN601937S, RSN601984S, RSN602002S, RSN602038S, RSN602211S, RSN602218S, RSN602275S, RSN602280S, RSN602458S, RSN602476S, RSN602497S, RSN602498S, RSN602500S, RSN600634S, RSN601473S, RSN601243S, RSN600187S, RSN600943S, RSN602058S, RSN600093S, RSN600102S, RSN600110S, RSN600933S, RSN600940S, RSN600962S, RSN601274S, RSN601769S, RSN602176S, RSN602578S, RSN602587S, RSN602921S, RSN601225S, RSN601298S, RSN600564S, RSN600405S, RSN602406S, RSN602732S, RSN600852S, RSN600878S, RSN601800S, RSN602168S, RSN600560S, RSN601462S, RSN601112S, RSN602674S, RSN601338S, RSN601727S, RSN602131S, RSN600667S, RSN601310S, RSN602248S, RSN600403S, RSN602636S, RSN600694S, RSN600935S, RSN601467S, RSN602163S, RSN602637S, RSN602639S, RSN602640S, RSN602642S, RSN601305S, RSN600973S, RSN601188S, RSN601189S, RSN601220S, RSN602288S, RSN602291S, RSN602293S, RSN602786S, RSN602525S, RSN602659S, RSN602661S, RSN602662S, RSN601464S, RSN602991S, RSN602875S, RSN602170S, RSN602418S, RSN602471S, RSN602518S, RSN602566S, RSN602879S, RSN602890S, RSN600867S, RSN600960S, RSN601218S, RSN601914S, RSN600620S, RSN601770S, RSN602225S, RSN602315S, RSN602364S, RSN600155S, RSN600165S, RSN600721S, RSN601952S, RSN602001S, RSN602757S, RSN600381S, RSN602289S, RSN600992S, RSN602281S, RSN602467S, RSN602546S, RSN602411S, RSN600143S, RSN602673S, RSN602228S, RSN602473S, RSN602480S, RSN602582S, RSN602583S, RSN602591S, RSN602595S, RSN602232S, RSN602233S, RSN602952S;  GTIN 00763000178383, Lot Serial Numbers:  RSN602082S, RSN602851S, RSN602638S, RSN602740S, RSN602741S, RSN602742S, RSN603203S, RSN603040S, RSN602779S, RSN602626S, RSN602814S, RSN603128S, RSN603046S, RSN601857S, RSN602603S, RSN603181S, RSN603213S, RSN601756S, RSN602597S, RSN600256S, RSN601222S, RSN601809S, RSN601774S, RSN601805S, RSN602623S, RSN602644S, RSN600263S, RSN601410S, RSN601713S, RSN602089S, RSN602672S, RSN602743S, RSN602922S, RSN603126S, RSN600073S, RSN600082S, RSN600275S, RSN600326S, RSN602075S, RSN602775S, RSN603173S, RSN601853S, RSN602099S, RSN602815S, RSN603115S, RSN601729S, RSN600085S, RSN600260S, RSN601781S, RSN602621S, RSN603158S, RSN601793S, RSN601824S, RSN601836S, RSN601714S, RSN603155S, RSN602620S, RSN602818S, RSN602981S, RSN603174S, RSN603198S, RSN600311S, RSN602630S, RSN601751S, RSN602622S, RSN600562S, RSN601370S, RSN602827S, RSN600245S, RSN601712S, RSN602097S, RSN601696S, RSN601803S, RSN601746S, RSN602846S, RSN601745S, RSN602095S, RSN602617S, RSN603092S, RSN601406S, RSN602837S, RSN602634S, RSN601127S, RSN601844S, RSN602625S, RSN603143S, RSN603192S, RSN603193S, RSN602831S, RSN602768S, RSN602770S, RSN603130S, RSN603135S, RSN600227S, RSN602841S, RSN601782S, RSN601738S, RSN602105S, RSN602629S, RSN602799S, RSN603075S, RSN602602S, RSN600323S, RSN601742S, RSN601743S, RSN602078S, RSN602090S, RSN602091S, RSN602103S, RSN600097S, RSN600259S, RSN601785S, RSN600078S, RSN601331S, RSN601344S, RSN602614S, RSN603079S, RSN603175S, RSN601859S, RSN602523S, RSN602609S, RSN602708S, RSN603153S, RSN603161S, RSN600089S, RSN601788S, RSN600228S, RSN602845S, RSN601849S, RSN600288S, RSN601757S, RSN602606S, RSN602847S, RSN600086S, RSN601779S, RSN601780S, RSN601775S, RSN602521S, RSN602563S, RSN602671S, RSN603022S, RSN603142S, RSN601765S, RSN601858S, RSN600014S, RSN600015S, RSN600087S, RSN600088S, RSN601865S, RSN601866S, RSN602615S, RSN602692S, RSN602773S, RSN602948S, RSN603036S, RSN603227S, RSN601728S, RSN601840S, RSN602322S, RSN601726S, RSN602085S, RSN601784S, RSN600021S, RSN600022S, RSN600023S, RSN600254S, RSN601815S, RSN601178S, RSN602819S, RSN601010S, RSN601697S, RSN602635S, RSN602735S, RSN602842S, RSN602861S, RSN600019S, RSN601792S, RSN603162S, RSN601789S, RSN602084S, RSN602631S, RSN602825S, RSN600229S, RSN602985S, RSN600081S, RSN602457S, RSN600232S, RSN600285S, RSN602667S, RSN600080S, RSN600286S, RSN602650S, RSN602849S, RSN603224S, RSN600287S, RSN602645S, RSN600289S, RSN600067S, RSN600101S, RSN600242S, RSN600261S, RSN600271S, RSN601744S, RSN601768S, RSN602080S, RSN602081S, RSN602600S, RSN602616S, RSN602733S, RSN602764S, RSN601814S, RSN600016S, RSN600084S, RSN602088S, RSN602098S, RSN602739S, RSN602632S, RSN602762S, RSN602763S, RSN602096S, RSN603039S, RSN603202S, RSN601826S, RSN602596S, RSN602598S, RSN602599S, RSN602618S, RSN602627S, RSN602628S, RSN602633S, RSN602641S, RSN602658S, RSN602736S, RSN602751S, RSN602766S, RSN602776S, RSN602809S, RSN602829S, RSN602839S, RSN602848S, RSN602852S, RSN602856S, RSN602987S, RSN603017S, RSN603035S, RSN603037S, RSN603042S, RSN603043S, RSN603044S, RSN603056S, RSN603059S, RSN603062S, RSN603080S, RSN603082S, RSN603084S, RSN603094S, RSN603118S, RSN603123S, RSN603137S, RSN603150S, RSN603151S, RSN603156S, RSN603164S, RSN603170S, RSN603172S, RSN603180S, RSN603182S, RSN603190S, RSN603194S, RSN603195S, RSN603197S, RSN603199S, RSN603200S, RSN603201S, RSN603204S, RSN603205S, RSN603209S, RSN603214S, RSN603226S, RSN603228S, RSN603144S, RSN603145S, RSN603146S, RSN602737S",
      "more_code_info": ""
    }
  ]
}