  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fairfield",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79873",
      "recalling_firm": "Datascope Corporation",
      "address_1": "15 Law Dr",
      "address_2": "N/A",
      "postal_code": "07004-3206",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.",
      "recall_number": "Z-1725-2018",
      "product_description": "NICU ECC Pack BEQ-TOP 39202, Custom Tubing Kit, Catalog No. 701067313R01",
      "product_quantity": "3",
      "reason_for_recall": "The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.",
      "recall_initiation_date": "20171019",
      "center_classification_date": "20180508",
      "termination_date": "20200424",
      "report_date": "20180516",
      "code_info": "Lot 3000061508"
    }
  ]
}