      "status": "Terminated",
      "city": "Fairfield",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79873",
      "recalling_firm": "Datascope Corporation",
      "address_1": "15 Law Dr",
      "address_2": "N/A",
      "postal_code": "07004-3206",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.",
      "recall_number": "Z-1723-2018",
      "product_description": "BEQ-T 25501 3/8\" Better Bladder Pack, Custom Tubing Kit, Catalog No. 709000370",
      "product_quantity": "10",
      "reason_for_recall": "The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.",
      "recall_initiation_date": "20171019",
      "center_classification_date": "20180508",
      "termination_date": "20200424",
      "report_date": "20180516",
      "code_info": "Lot 3000060037"
    }
  ]
}