  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.",
      "address_2": "",
      "product_quantity": "345",
      "code_info": "Material numbers: 10165977, 10045692 , and 10165888  with serial numbers: 32187  31158  31215  31222  31313  31354  32118  32120  32123  32125  32126  32127  32129  32139  29161  27229  29185  27415  27212  29364  29186  29165  29233  29182  27235  29338  29176  29376  27232  29181  27369  29295  27230  29198  27244  29286  29126  27338  29302  27231  27334  29394  29264    32141  32156  32159",
      "center_classification_date": "20140604",
      "distribution_pattern": "US Nationwide Distribution in the states of IA, TX, FL, NE, NY, OH, NC, KS, IL, CA, OK including PR.",
      "state": "PA",
      "product_description": "Siemens SOMATOM Emotion 16, Emotion 6, and SOMATOM Spirit Computed Tomography Systems     Product Usage:  To produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1721-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67967",
      "termination_date": "20150316",
      "more_code_info": "",
      "recall_initiation_date": "20140321",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}