{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98598",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).",
      "recall_number": "Z-1719-2026",
      "product_description": "Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows:    1. NON VASCULAR PACK, Medline SKU # DYNJ68037C;   2. VEIN, Medline SKU # DYNJ89845;   3. VEIN, Medline SKU # DYNJ89845A.",
      "product_quantity": "2980 kits",
      "reason_for_recall": "Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.",
      "recall_initiation_date": "20260227",
      "center_classification_date": "20260407",
      "report_date": "20260415",
      "code_info": "Medline SKU # DYNJ68037C, UDI/DI each 10198459239328, UDI/DI case 40198459239329, Lot Number: 25HBV397;  Medline SKU # DYNJ89845, UDI/DI each 10198459292460, UDI/DI case 40198459292461, Lot Number: 25CBU222;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26BBC195;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26ABI611;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25LBK402;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25KBH371;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25IBK975;  Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25HBQ426.",
      "more_code_info": ""
    }
  ]
}