      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "1266 Kifer Rd Bldg 100",
      "reason_for_recall": "Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.",
      "address_2": "",
      "product_quantity": "2",
      "code_info": "lot 380652-20",
      "center_classification_date": "20140604",
      "distribution_pattern": "Medical facilities in TX and FL",
      "state": "CA",
      "product_description": "Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000.  The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Intuitive Surgical, Inc.",
      "recall_number": "Z-1718-2014",
      "initial_firm_notification": "Other",
      "product_type": "Devices",
      "event_id": "68358",
      "termination_date": "20140604",
      "more_code_info": "",
      "recall_initiation_date": "20140509",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}