{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mundelein",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84759",
      "recalling_firm": "Sysmex America, Inc.",
      "address_1": "1 Sysmex Way",
      "address_2": "",
      "postal_code": "60060-9528",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA: FL, VT, IL.",
      "recall_number": "Z-1717-2020",
      "product_description": "Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.",
      "product_quantity": "40 units",
      "reason_for_recall": "When excessive amounts of urine spills on  the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.",
      "recall_initiation_date": "20200121",
      "center_classification_date": "20200415",
      "termination_date": "20230130",
      "report_date": "20200422",
      "code_info": "Serial Numbers:  11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967, 14968, 14969, 14970, 14972, 14973, 14974, 14975, 14976, 14977, 14978, 14979, 15017, 15018, 15619, 15620",
      "more_code_info": ""
    }
  ]
}