      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71209",
      "recalling_firm": "Advanced Sterilization Products",
      "address_1": "33 Technology Dr",
      "address_2": "N/A",
      "postal_code": "92618-2346",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  Australia, Singapore, Canada, and Thailand.",
      "recall_number": "Z-1715-2015",
      "product_description": "EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002.    The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.",
      "product_quantity": "44 units",
      "reason_for_recall": "Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.",
      "recall_initiation_date": "20150511",
      "center_classification_date": "20150605",
      "termination_date": "20160706",
      "report_date": "20150617",
      "code_info": "5040207003  5040207004  5040308003  5042100008  5042100016  5042100017  5042100018  5042100019  5042100020  5042100021  5042100022  5042100023  5042100024  5042100025  5042100033  5042100034  5042100035  5042100036  5042100037  5042100038  5042100039  5042100040  5042100060  5042100061  5042100063  5042100064  5042100065  5042100066  5042100067  5042110002  5042110003  5042110004  5042110005  5042110006  5042120002  5042120006  5042130009  5042130011  5042140029  5042140030  5042140031  5042140032  5042140038  5042140039"
    }
  ]
}