      "country": "United States",
      "city": "Mansfield",
      "address_1": "120 Forbes Blvd Ste 170",
      "reason_for_recall": "The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered.  There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.  Error 115 can also occur if dendrites form or if the pump is dropped.",
      "address_2": "",
      "product_quantity": "809",
      "code_info": "Serial Numbers:  10AX4A03  10BY4A40  10EY4A35  10GX4A83  10HW4A31  10HY4C11  10JW4A28  10KW4A42  10KX4A48  10AX4A04  10BY4A41  10EY4A36  10GX4A84  10HW4A45  10HY4C12  10JW4A29  10KW4A43  10KX4A49  10AX4A06  10BY4A42  10EY4A37  10GX4A85  10HW4A57  10HY4C13  10JW4A30  10KW4A44  10KX4A50  10AX4A10  10BY4A44  10EY4A38  10GX4A86  10HY4A77  10HY4C15  10JW4A31  10KW4A45  10KX4A52  10AX4A12  10BY4A45  10EY4A40  10GX4A87  10HY4A78  10HY4C16  10JW4A32  10KW4A46  10KX4A56  10AX4A14  10BY4A46  10EY4A41  10GX4A88  10HY4A79  10HY4C17  10JW4A33  10KW4A47  10KX4A57  10AX4A21  10BY4A47  10EY4A43  10GX4A89  10HY4A80  10HY4C18  10JW4A34  10KW4A48  10KX4A58  10AX4A22  10BY4A48  10EY4A44  10GX4A92  10HY4A81  10HY4C19  10JW4A35  10KW4A49  10KX4A59  10AX4A23  10BY4A49  10EY4A45  10GX4A94  10HY4A82  10HY4C20  10JW4A36  10KW4A50  10KX4A60  10AX4A28  10BY4A50  10EY4A46  10GX4A96  10HY4A83  10HY4C22  10JW4A37  10KW4A51  10KX4A62  10AX4A30  10BY4A51  10EY4A47  10GX4A97  10HY4A84  10HY4C23  10JW4A38  10KW4A52  10KX4A63  10AX4A36  10BY4A52  10EY4A48  10GX4A98  10HY4A85  10HY4C26  10JW4A39  10KW4A53  10KX4A64  10AX4A37  10BY4A53  10EY4A51  10GX4A99  10HY4A87  10HY4C27  10JW4A40  10KW4A54  10KX4A65  10AX4A39  10BY4A54  10EY4A52  10GX4B02  10HY4A90  10HY4C28  10JW4A41  10KW4A55  10KX4A66  10AX4A42  10BY4A55  10EY4A53  10GX4B03  10HY4A91  10HY4C29  10JW4A42  10KW4A56  10KX4A67  10AX4A48  10BY4A56  10EY4A54  10GX4B04  10HY4A92  10HY4C30  10JW4A43  10KW4A57  10KX4A68  10AX4A51  10BY4A57  10EY4A55  10GX4B05  10HY4A95  10HY4C32  10JW4A44  10KW4A58  10KX4A69  10AX4A73  10BY4A58  10EY4A56  10GX4B06  10HY4A97  10HY4C34  10JW4A45  10KW4A59  10KX4A70  10AX4A74  10BY4A59  10EY4A58  10GX4B07  10HY4A98  10HY4C35  10JW4A46  10KW4A60  10KX4A71  10AX4A76  10BY4A60  10EY4A59  10GX4B08  10HY4B01  10HY4C36  10JW4A47  10KW4A61  10KX4A72  10AX4A81  10BZ4A24  10EY4A60  10GX4B09  10HY4B02  10HY4C37  10JW4A48  10KW4A63  10KX4A73  10AX4A83  10BZ4A26  10EY4A61  10GX4B10  10HY4B03  10HY4C39  10JY4A01  10KW4A64  10KX4A74  10AX4A85  10BZ4A32  10EY4A62  10GX4B11  10HY4B05  10HY4C40  10JY4A03  10KW4A65  10KX4A76  10AX4A87  10BZ4A46  10EY4A63  10GX4B12  10HY4B07  10HY4C41  10JY4A06  10KW4A66  10KX4A77  10AX4A88  10BZ4A70  10EY4A64  10GX4B13  10HY4B08  10HY4C42  10JY4A07  10KW4A67  10KX4A78  10AX4A89  10BZ4A71  10EY4A65  10GX4B14  10HY4B10  10HY4C46  10JY4A09  10KW4A68  10KX4A79  10AX4A93  10CZ4A02  10EY4A66  10GX4B15  10HY4B12  10HY4C47  10JY4A10  10KW4A70  10KX4A80  10AX4A95  10CZ4A03  10EY4A67  10GX4B16  10HY4B16  10IY2A05  10JY4A15  10KW4A71  10KX4A81  10AX4A96  10CZ4A04  10EY4A69  10GX4B21  10HY4B17  10IY2A26  10JY4A16  10KW4A72  10KX4A82  10AX4A98  10DX4A03  10EY4A70  10GX4B22  10HY4B18  10IY4A01  10JY4A18  10KW4A73  10KX4A85  10AX4B06  10DX4A07  10EY4A72  10GX4B25  10HY4B19  10IY4A02  10JY4A20  10KW4A74  10KY4A02  10AX4B08  10DX4A09  10EY4A73  10GY4A02  10HY4B20  10IY4A03  10JY4A21  10KW4A75  10KY4A03  10AX4B09  10DX4A10  10EY4A74  10GY4A03  10HY4B22  10IY4A04  10JY4A25  10KW4A76  10KY4A04  10AX4B16  10DX4A13  10EY4A75  10GY4A05  10HY4B23  10IY4A05  10JY4A28  10KW4A77  10KY4A06  10AX4B17  10DX4A21  10EY4A76  10GY4A06  10HY4B24  10IY4A09  10JY4A29  10KW4A78  10KY4A07  10DX4A24  10EY4A81  10GY4A11  10HY4B26  10IY4A11  10KV4A01  10KW4A80  10KY4A10  10AX4B26  10DX4A32  10EY4A82  10GY4A12  10HY4B27  10IY4A13  10KW4A01  10KW4A81  10KY4A11  10AX4B27  10DX4A33  10EY4A83  10GY4A13  10HY4B28  10IY4A15  10KW4A02  10KW4A83  10KY4A12  10AX4B29  10DX4A34  10EY4A87  10GY4A14  10HY4B29  10IY4A17  10KW4A03  10KW4A84  10KY4A13  10AX4B53  10DX4A41  10EY4A88  10GY4A15  10HY4B30  10IY4A19  10KW4A04  10KW4A85  10KY4A15  10AX4B78  10DX4A44  10EY4A89  10GY4A16  10HY4B33  10IY4A25  10KW4A05  10KW4A86  10KY4A17  10AX4B79  10DX4A48  10EY4A90  10GY4A18  10HY4B34  10IY4A26  10KW4A06  10KW4A87  10KY4A18  10AX4B80  10DX4A52  10EY4A92  10GY4A19  10HY4B35  10IY4A27  10KW4A07  10KW4A88  10KY4A20  10AX4B81  10DX4A60  10EY4A93  10GY4A20  10HY4B36  10IY4A28  10KW4A08  10KW4A89  10KY4A21  10AX4B82  10DX4A62  10EY4A94  10GY4A21  10HY4B37  10IY4A29  10KW4A09  10KW4A90  10KY4A25  10AX4B84  10DX4A65  10EY4A95  10GY4A22  10HY4B40  10IY4A31  10KW4A10  10KW4A91  10KY4A26  10AX4B85  10DX4A67  10EY4A96  10GY4A23  10HY4B41  10IY4A33  10KW4A101  10KW4A92  10KY4A27  10AX4B86  10DX4A71  10EY4A97  10GY4A25  10HY4B42  10IY4A37  10KW4A103  10KW4A93  10KY4A30  10AX4B88  10DX4A74  10EY4A98  10GY4A26  10HY4B43  10IY4A38  10KW4A104  10KW4A94  10KY4A31  10AX4B90  10DX4A79  10EY4A99  10GY4A27  10HY4B44  10IY4A41  10KW4A105  10KW4A95  10KY4A33  10AX4B91  10DX4A82  10GX4A03  10GY4A28  10HY4B45  10IY4A43  10KW4A106  10KW4A96  10KY4A35  10AX4B92  10DX4A85  10GX4A04  10GY4A30  10HY4B48  10IY4A44  10KW4A108  10KW4A98  10KY4A36  10AX4B94  10DX4A89  10GX4A08  10GY4A31  10HY4B49  10IY4A46  10KW4A109  10KW4A99  10KY4A37  10AX4B96  10DX4A95  10GX4A09  10GY4A32  10HY4B50  10IY4A50  10KW4A11  10KX4A01  10KY4A39  10AX4B97  10DX4A96  10GX4A10  10GY4A34  10HY4B51  10IY4A51  10KW4A110  10KX4A02  10KY4A40  10AZ4A08  10DX4A98  10GX4A12  10GY4A35  10HY4B53  10IY4A52  10KW4A111  10KX4A04  10KY4A41  10AZ4A10  10DX4A99  10GX4A14  10GY4A36  10HY4B54  10IY4A54  10KW4A113  10KX4A05  10KY4A43  10AZ4A12  10DX4B02  10GX4A15  10GY4A39  10HY4B55  10IY4A56  10KW4A116  10KX4A06  10KY4A44  10AZ4A14  10DX4B03  10GX4A16  10GY4A40  10HY4B56  10IY4A58  10KW4A117  10KX4A07  10KY4A45  10AZ4A15  10DX4B07  10GX4A19  10GY4A41  10HY4B62  10IY4A59  10KW4A119  10KX4A08  10KY4A47  10AZ4A18  10DX4B15  10GX4A21  10GY4A42  10HY4B63  10IY4A60  10KW4A12  10KX4A09  10KY4A50  10AZ4A19  10DX4B17  10GX4A30  10GY4A44  10HY4B65  10IY4A61  10KW4A120  10KX4A10  10KY4A51  10AZ4A22  10DX4B21  10GX4A31  10GY4A45  10HY4B66  10IY4A62  10KW4A121  10KX4A11  10KY4A60  10AZ4A24  10DX4B25  10GX4A32  10GY4A47  10HY4B69  10JW4A01  10KW4A122  10KX4A12  10KY4A61  10AZ4A31  10DX4B26  10GX4A33  10GY4A49  10HY4B70  10JW4A02  10KW4A123  10KX4A13  10KY4A62  10AZ4A33  10EY4A01  10GX4A34  10GY4A50  10HY4B74  10JW4A03  10KW4A125  10KX4A14  10KY4A63  10AZ4A43  10EY4A02  10GX4A37  10GY4A51  10HY4B76  10JW4A04  10KW4A13  10KX4A15  10KY4A64  10AZ4A54  10EY4A03  10GX4A38  10GY4A52  10HY4B78  10JW4A05  10KW4A14  10KX4A16  10KY4A65  10AZ4A57  10EY4A04  10GX4A39  10GY4A53  10HY4B79  10JW4A06  10KW4A15  10KX4A17  10KY4A67  10AZ4A59  10EY4A05  10GX4A40  10GY4A54  10HY4B80  10JW4A07  10KW4A16  10KX4A18  10KY4A68  10AZ4A61  10EY4A08  10GX4A44  10GY4A55  10HY4B81  10JW4A08  10KW4A17  10KX4A19  10KY4A69  10AZ4A65  10EY4A11  10GX4A45  10GY4A56  10HY4B83  10JW4A09  10KW4A18  10KX4A20  10KY4A71  10AZ4A70  10EY4A13  10GX4A46  10GY4A57  10HY4B84  10JW4A10  10KW4A19  10KX4A21  10KY4A72  10AZ4A73  10EY4A14  10GX4A47  10GY4A58  10HY4B85  10JW4A11  10KW4A20  10KX4A22  10KY4A75  10BY4A25  10EY4A15  10GX4A48  10GY4A59  10HY4B86  10JW4A12  10KW4A21  10KX4A25  10KY4A78  10BY4A26  10EY4A16  10GX4A57  10GY4A60  10HY4B90  10JW4A13  10KW4A23  10KX4A26  10KY4A79  10BY4A27  10EY4A17  10GX4A62  10GY4A61  10HY4B91  10JW4A14  10KW4A24  10KX4A27  10KY4A82  10GX4A65  10GY4A63  10HY4B96  10JW4A16  10KW4A26  10KX4A29  10KY4A84  10BY4A30  10EY4A21  10GX4A67  10GY4A64  10HY4B97  10JW4A17  10KW4A27  10KX4A30  10KY4A86  10BY4A31  10EY4A22  10GX4A69  10GY4A65  10HY4B99  10JW4A18  10KW4A29  10KX4A32  10KY4A87  10BY4A32  10EY4A24  10GX4A70  10GY4A66  10HY4C01  10JW4A19  10KW4A30  10KX4A39  10KY4A90  10BY4A33  10EY4A25  10GX4A71  10GY4A67  10HY4C02  10JW4A20  10KW4A31  10KX4A40  10KY4A91  10BY4A34  10EY4A26  10GX4A72  10GY4A70  10HY4C03  10JW4A21  10KW4A32  10KX4A41  10KY4A92  10BY4A35  10EY4A27  10GX4A73  10GY4A71  10HY4C04  10JW4A22  10KW4A35  10KX4A42  10KY4A93  10BY4A36  10EY4A28  10GX4A74  10GY4A72  10HY4C06  10JW4A24  10KW4A38  10KX4A43  10KY4A94  10BY4A37  10EY4A29  10GX4A76  10GY4A73  10HY4C07  10JW4A25  10KW4A39  10KX4A44  10KY4A97  10BY4A38  10EY4A31  10GX4A81  10GY4A74  10HY4C08  10JW4A26  10KW4A40  10KX4A46  10KY4A99  10BY4A39  10EY4A33  10GX4A82  10HW4A29  10HY4C09  10JW4A27  10KW4A41  10KX4A47",
      "center_classification_date": "20190603",
      "distribution_pattern": "US nationwide distribution.",
      "state": "MA",
      "product_description": "Prometra Programmable Pump, Cat. No. 11827   For intrathecal infusion.",
      "report_date": "20190612",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Flowonix Medical Inc",
      "recall_number": "Z-1714-2019",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "82586",
      "termination_date": "20200605",
      "more_code_info": "",
      "recall_initiation_date": "20190321",
      "postal_code": "02048-1150",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}