  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73575",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "**Domestic** AR, AZ, CA, CT, CO, DE, FL, GA, IL, LA, MI, MD, MS, MA, NY, NV, NJ, OH, OR, PA, NC, RI, TN, WA & Washington D.C; **Internationally** Algeria, Australia, Austria, Belgium, Canada, China, France, French Polynesia, Germany, Iran, Italy, Japan, Netherlands, New Zealand, Norway, Republic of Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom",
      "recall_number": "Z-1712-2016",
      "product_description": "RayStation Therapy Treatment Planning System Stand-alone Software 3.0, 3.5, 4.0, 4.5 and 4.7., designed for treatment planning and analysis of radiation therapy.",
      "product_quantity": "1,264 Total units (552 units domestically & 711 units internationally)",
      "reason_for_recall": "A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right clicking, pressing Ctrl-S or Ctrl-Z while the left mouse button is held down. This bug does not affect dose computations, which are based on the stored system state.",
      "recall_initiation_date": "20160210",
      "center_classification_date": "20160524",
      "termination_date": "20170424",
      "report_date": "20160601",
      "code_info": "3.0.0.251\u001d, \u001c3.5.0.16\u001d, \u001c3.5.1.6\u001d, \u001c4.0.0.14\u001d, \u001c4.0.1.4\u001d, \u001c4.0.2.9\u001d, \u001c4.0.3.4\u001d, \u001c4.5.0.19\u001d, \u001c4.5.1.14\u001d, \u001c4.5.2.7\u001d, \u001c4.7.0.15\u001d, \u001c4.7.1.10\u001d, \u001c4.7.2.5\u001d, \u001c4.7.3.13\u001d or \u001c4.7.4.4"
    }
  ]
}