      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76688",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe",
      "recall_number": "Z-1711-2017",
      "product_description": "UniCel DxC 600i SYNCHRON Access Clinical System, Part No. A25637, A25638    Product Usage:  The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.",
      "product_quantity": "851 units total for all products (291 units in US)",
      "reason_for_recall": "Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.",
      "recall_initiation_date": "20170307",
      "center_classification_date": "20170404",
      "termination_date": "20180301",
      "report_date": "20170412",
      "code_info": "Serial No. (Part No. A25637) 800999,801019    Serial No. (Part No. A25638)800995,800996,800997,800998,801000,801001,801002,801003,801004,  801005,801006,801007,801008,801009,801010,801011,801012,801013,  801014,801015,801016,801017,801018,801020,801021,801022,801023,  801024,801025,801026,801027"
    }
  ]
}