{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76688",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe",
      "recall_number": "Z-1710-2017",
      "product_description": "Access 2 assay Immunoassay System, Remanufactured, Part No. 386220    Product Usage:  The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.",
      "product_quantity": "851 units total for all products (291 units in US)",
      "reason_for_recall": "Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.",
      "recall_initiation_date": "20170307",
      "center_classification_date": "20170404",
      "termination_date": "20180301",
      "report_date": "20170412",
      "code_info": "Serial No. 501012,501110,501190,501224,501294,501371,501394,501403,501440,  501443,501490,501501,501673,501681,501693,501861,501920,501926,  501936,501970,502165,502196,502319,502331,502332,502368,502458,  502475,502506,502543,502549,502556,502574,502640,502647,502651,  502791,502843,502848,503153,503279,503313,503354,503384,503392,  503424,503543,503627,503678,504046,504071,504072,504092,504112,  504195,504322,504365,504370,504756,504767,504902,505042,505195,  505251,505254,505644,505795,505840,505947,506002,506026,506108,  506119,506153,506187,506371,506406,506560,506586,506664,506708,  506766,506786,506837,506846,507039,507120,507161,507162,507214,  507229,507248,507282,507353,507397,507480,507511,507604,507619,  507842,507929,508019,508096,508179,508227,508313,508334,508412, 508478,508509,508516,508767,509051,509087,509113,509164,509253,  509297,509632,509837,510049,511042,511047,511180,511383,700024,  800321,800341,800443,800608,800611,800650,800849"
    }
  ]
}