  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73945",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of: AE, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI",
      "recall_number": "Z-1710-2016",
      "product_description": "3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile    Product Usage:  The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).",
      "product_quantity": "1274 units",
      "reason_for_recall": "The sterile packaging may be compromised.  A protective cap is to be secured on the tip of the drill bit.  If the protective cap is not secured, the drill bit may puncture the sterile barrier.",
      "recall_initiation_date": "20160418",
      "center_classification_date": "20160523",
      "termination_date": "20170202",
      "report_date": "20160601",
      "code_info": "Catalog ID 03.010.100S  Lot # 5525872, 6153368, 6530901, 5548707, 6197315, 6581051  5575918, 6199232, 6676406, 5603463, 6231850, 6721724  5649117, 6256560, 6734876, 5668880, 6305842, 6808463  5724639, 6319050, 6831468, 5742276, 6320150, 6997294  5807341, 6343980, 7039877, 6012770, 6401819, 7052565  6045597, 6448707, 7081479, 6057914, 6472100, 7378832  6108342, 6501856, 7410311, 6142443, 6503351, 7466207  7554670, 7620971, 7675571, 7711694, 7801251, 7918026  9815103, 9828932, 9905850, 9939769, H033348"
    }
  ]
}