  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Baltimore",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76660",
      "recalling_firm": "Philips Visicu",
      "address_1": "217 E Redwood St Ste 1900",
      "address_2": "N/A",
      "postal_code": "21202-3315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US Nationwide",
      "recall_number": "Z-1708-2017",
      "product_description": "eCare Coordinator    Product Usage:   is software intended for use in data aggregation, patient interaction facilitation, storage and clinical information management with independent physiological devices and ancillary systems that are connected either directly or through networks.  The software is intended to provide patient information from the patient location through networking technology to a remote care team. eCare Coordinator does not send any real time alarms and is not intended to provide automated treatment decisions. This software is an informational tool only and is not to be used as a substitute for professional judgment of healthcare providers in diagnosing and treating patients.",
      "product_quantity": "26",
      "reason_for_recall": "eCareCoordinator (eCC) is intended for use in data aggregation, patient interaction facilitation, storage and clinical information management with independent  physiological devices and ancillary systems that are connected either directly or through networks.  The software is intended to provide patient information from the patient location through networking technology to a remote care team.    A software defect was discovered in the Philips eCareCoordinator (eCC) Clinical User Interface that can at times cause missing or redundant data to be saved without notification to the user.",
      "recall_initiation_date": "20170302",
      "center_classification_date": "20170404",
      "termination_date": "20171027",
      "report_date": "20170412",
      "code_info": "453564506091   eCareCoordinator"
    }
  ]
}