{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92328",
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution",
      "recall_number": "Z-1707-2023",
      "product_description": "ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1;  Implantable Cardioverter Defibrillators",
      "product_quantity": "2282 units",
      "reason_for_recall": "There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.",
      "recall_initiation_date": "20230510",
      "center_classification_date": "20230616",
      "report_date": "20230628",
      "code_info": "GTIN 00643169217911, Lot Serial Numbers:  BWE613480S, BWE613110S, BWE613432S, BWE613202S, BWE613100S, BWE613096S, BWE613185S, BWE613258S, BWE613431S, BWE612802S, BWE612865S, BWE613214S, BWE613235S, BWE612884S, BWE612956S, BWE613098S, BWE613122S, BWE613143S, BWE613171S, BWE613212S, BWE613216S, BWE613251S, BWE613470S, BWE613162S, BWE613238S, BWE613459S, BWE613430S, BWE613153S, BWE613484S, BWE613168S, BWE612948S, BWE612950S, BWE612970S, BWE613487S, BWE613090S, BWE613224S, BWE613180S, BWE613077S, BWE612815S, BWE612940S, BWE613086S, BWE612721S, BWE613194S, BWE612880S, BWE613125S, BWE613221S, BWE612822S, BWE612985S, BWE612928S, BWE613446S, BWE612939S, BWE612837S, BWE613097S, BWE613095S, BWE613104S, BWE613211S, BWE613463S, BWE613460S, BWE612679S, BWE612960S, BWE613149S, BWE613150S, BWE613247S, BWE612981S, BWE612987S, BWE613145S, BWE613147S, BWE613485S, BWE612986S, BWE613443S, BWE612844S, BWE612686S, BWE612824S, BWE612853S, BWE612923S, BWE612976S, BWE612935S, BWE612982S, BWE612993S, BWE613155S, BWE613124S, BWE612930S, BWE612949S, BWE612833S, BWE613115S, BWE613182S, BWE613121S, BWE612922S, BWE612990S, BWE613197S, BWE612801S, BWE612988S, BWE613111S, BWE613116S, BWE613146S, BWE613183S, BWE613352S, BWE613141S, BWE613466S, BWE612675S, BWE612785S, BWE612796S, BWE612961S, BWE613089S, BWE613101S, BWE613250S, BWE613253S, BWE613448S, BWE613467S, BWE612978S, BWE612685S, BWE612974S, BWE612876S, BWE612917S, BWE612977S, BWE613082S, BWE613102S, BWE613156S, BWE613441S, BWE613444S, BWE613473S, BWE613172S, BWE612791S, BWE612792S, BWE612793S, BWE612794S, BWE612795S, BWE612797S, BWE612798S, BWE612799S, BWE612800S, BWE612893S, BWE612900S, BWE612954S, BWE612957S, BWE613259S, BWE612953S, BWE613181S, BWE613472S, BWE613195S, BWE613468S, BWE613433S, BWE613084S, BWE613201S, BWE613469S, BWE612827S, BWE613088S, BWE613184S, BWE612963S, BWE612983S, BWE613126S, BWE613215S, BWE613445S, BWE613471S, BWE612823S, BWE612840S, BWE612932S, BWE612885S, BWE612769S, BWE613196S, BWE613140S, BWE613229S, BWE612779S, BWE613239S, BWE613163S, BWE613236S, BWE612836S, BWE612899S, BWE613245S, BWE612776S, BWE612851S, BWE613170S, BWE612847S, BWE612965S, BWE613240S, BWE612966S, BWE612971S, BWE613187S, BWE613243S, BWE612894S, BWE612955S, BWE613081S, BWE613119S, BWE613160S, BWE613204S, BWE612770S, BWE612832S, BWE612423S, BWE612972S, BWE612984S, BWE613087S, BWE613148S, BWE613152S, BWE613482S, BWE612980S, BWE613228S, BWE613449S, BWE613434S, BWE612929S, BWE612938S, BWE612898S, BWE612811S, BWE612882S, BWE613161S, BWE612818S, BWE613112S, BWE612842S, BWE612806S, BWE612821S, BWE612969S, BWE613123S, BWE613157S, BWE613158S, BWE612825S, BWE613128S, BWE612879S, BWE612889S, BWE612967S, BWE613207S, BWE612817S, BWE613085S, BWE613474S, BWE613129S, BWE612979S, BWE613210S, BWE613462S, BWE613252S, BWE613457S, BWE612908S, BWE612909S, BWE612914S, BWE612915S, BWE612919S, BWE613002S, BWE613005S, BWE613006S, BWE613008S, BWE613017S, BWE613020S, BWE613026S, BWE613113S, BWE613114S, BWE613118S, BWE613127S, BWE613173S, BWE613174S, BWE613175S, BWE613176S, BWE613178S, BWE613455S, BWE613465S, BWE613475S, BWE613476S, BWE613477S, BWE613478S, BWE613479S, BWE613481S, BWE613483S, BWE612772S, BWE613131S, BWE613093S, BWE612886S, BWE612887S, BWE612901S, BWE613244S, BWE613256S, BWE612947S, BWE613199S, BWE612422S, BWE612911S, BWE613151S, BWE613154S, BWE613219S, BWE613454S, BWE613190S, BWE612677S, BWE612924S, BWE613189S, BWE612975S, BWE613220S, BWE612888S, BWE613142S, BWE613451S, BWE613208S, BWE612958S, BWE612814S, BWE612789S, BWE613120S, BWE613188S, BWE612712S, BWE613106S, BWE613164S, BWE613186S, BWE613254S, BWE612816S, BWE613169S, BWE613218S, BWE613231S, BWE613233S, BWE612778S, BWE612921S, BWE613166S, BWE612998S, BWE613001S, BWE613144S, BWE613078S, BWE613107S, BWE613130S, BWE613222S, BWE613223S, BWE613461S, BWE612944S, BWE613234S, BWE613117S, BWE613486S, BWE613159S, BWE613179S, BWE612936S, BWE612920S, BWE613203S, BWE613248S, BWE612881S, BWE613103S, BWE613192S, BWE612807S, BWE613209S, BWE612910S, BWE613191S;  GTIN:  00643169969568, Lot Serial Numbers:  BWE612952S, BWE613091S, BWE613094S, BWE613297S, BWE613299S, BWE613316S, BWE613326S, BWE613429S, BWE613589S, BWE613591S, BWE613597S, BWE613659S, BWE613664S, BWE613775S, BWE612704S, BWE612717S, BWE613092S, BWE613138S, BWE613343S, BWE612681S, BWE612692S, BWE613007S, BWE612666S, BWE612667S, BWE612687S, BWE612647S, BWE612803S, BWE613232S, BWE613307S, BWE613490S, BWE612646S, BWE613285S, BWE612732S, BWE613217S, BWE613237S, BWE613287S, BWE613289S, BWE613300S, BWE613344S, BWE613489S, BWE613301S, BWE613165S, BWE613246S, BWE613206S, BWE613495S, BWE613598S, BWE613227S, BWE613230S, BWE612658S, BWE613497S, BWE612657S, BWE613354S, BWE612730S, BWE613269S, BWE613588S, BWE612690S, BWE613132S, BWE612782S, BWE613341S, BWE612421S, BWE612734S, BWE612695S, BWE612731S, BWE612933S, BWE613257S, BWE613488S, BWE612683S, BWE612688S, BWE612662S, BWE612670S, BWE612698S, BWE612703S, BWE612705S, BWE612838S, BWE613996S, BWE612653S, BWE612684S, BWE613313S, BWE612661S, BWE612663S, BWE612665S, BWE612672S, BWE612693S, BWE612697S, BWE612708S, BWE612709S, BWE612716S, BWE612718S, BWE612719S, BWE612722S, BWE612723S, BWE612724S, BWE613260S, BWE613308S, BWE613498S, BWE612783S, BWE613262S, BWE612962S, BWE613304S, BWE613492S, BWE612864S, BWE612926S, BWE612927S, BWE612649S, BWE612650S, BWE612651S, BWE613364S, BWE613610S, BWE613213S, BWE613275S, BWE613616S, BWE612896S, BWE612906S, BWE613019S, BWE613249S, BWE613295S, BWE612867S, BWE612868S, BWE612870S, BWE612754S, BWE612843S, BWE612855S, BWE612856S, BWE612857S, BWE612858S, BWE612859S, BWE612861S, BWE612862S, BWE613317S, BWE613324S, BWE613099S, BWE613226S, BWE613242S, BWE613511S, BWE613613S, BWE612804S, BWE613255S, BWE612854S, BWE612660S, BWE612714S, BWE612715S, BWE612720S, BWE612746S, BWE612750S, BWE612784S, BWE612831S, BWE612835S, BWE612959S, BWE613290S, BWE613296S, BWE613302S, BWE613320S, BWE613321S, BWE612895S, BWE613310S, BWE613312S, BWE613561S, BWE612737S, BWE612786S, BWE612973S, BWE612991S, BWE612994S, BWE612996S, BWE613010S, BWE613012S, BWE613013S, BWE613014S, BWE613016S, BWE613018S, BWE613034S, BWE613074S, BWE613345S, BWE613389S, BWE613404S, BWE613426S, BWE613442S, BWE613453S, BWE613539S, BWE613544S, BWE613545S, BWE613549S, BWE613563S, BWE613572S, BWE613579S, BWE613580S, BWE613581S, BWE613620S, BWE613631S, BWE613655S, BWE613704S, BWE613705S, BWE613713S, BWE613721S, BWE613758S, BWE613773S, BWE613774S, BWE613776S, BWE613777S, BWE613788S, BWE613797S, BWE613804S, BWE613805S, BWE613806S, BWE613807S, BWE613811S, BWE613812S, BWE613813S, BWE613814S, BWE613815S, BWE613816S, BWE613817S, BWE613818S, BWE613836S, BWE613837S, BWE613843S, BWE613844S, BWE613846S, BWE613847S, BWE613921S, BWE613922S, BWE613923S, BWE613435S, BWE612997S, BWE612748S, BWE612751S, BWE613205S, BWE613761S, BWE613819S, BWE613820S, BWE613831S, BWE612829S, BWE612839S, BWE612849S, BWE612866S, BWE612964S, BWE613021S, BWE613022S, BWE613039S, BWE613044S, BWE613053S, BWE613055S, BWE613365S, BWE613409S, BWE613413S, BWE613427S, BWE613437S, BWE613447S, BWE613450S, BWE613502S, BWE613560S, BWE613562S, BWE613633S, BWE613999S, BWE614000S, BWE612897S, BWE612902S, BWE612912S, BWE612913S, BWE612916S, BWE612943S, BWE612951S, BWE613023S, BWE613024S, BWE613033S, BWE613045S, BWE613065S, BWE612761S, BWE612790S, BWE612808S, BWE612848S, BWE613496S, BWE613565S, BWE613566S, BWE613629S, BWE613654S, BWE613841S, BWE613904S, BWE613907S, BWE613060S, BWE613061S, BWE613062S, BWE613064S, BWE613066S, BWE613068S, BWE613069S, BWE613071S, BWE612738S, BWE612745S, BWE612747S, BWE612760S, BWE612766S, BWE612812S, BWE612873S, BWE612877S, BWE612892S, BWE612905S, BWE613029S, BWE613035S, BWE613041S, BWE613042S, BWE613050S, BWE613051S, BWE613057S, BWE613058S, BWE613059S, BWE613063S, BWE613067S, BWE613070S, BWE613298S, BWE613311S, BWE613315S, BWE613327S, BWE613329S, BWE612739S, BWE612741S, BWE612749S, BWE613436S, BWE613725S, BWE613759S, BWE613799S, BWE613984S, BWE613997S, BWE612874S, BWE612883S, BWE612918S, BWE612945S, BWE612989S, BWE613000S, BWE613025S, BWE613040S, BWE613073S, BWE613076S, BWE613139S, BWE613198S, BWE613279S, BWE613283S, BWE613370S, BWE613384S, BWE613391S, BWE613392S, BWE613397S, BWE613412S, BWE613425S, BWE613439S, BWE613452S, BWE613458S, BWE613494S, BWE613503S, BWE613506S, BWE613507S, BWE613541S, BWE613542S, BWE613543S, BWE613574S, BWE613621S, BWE613630S, BWE613656S, BWE613662S, BWE613663S, BWE613702S, BWE613711S, BWE613765S, BWE613794S, BWE613801S, BWE613839S, BWE613906S, BWE613985S, BWE613992S, BWE613993S, BWE613998S, BWE614001S, BWE613072S, BWE613382S, BWE613513S, BWE613564S, BWE613724S, BWE613827S, BWE614003S, BWE612742S, BWE613810S, BWE613910S, BWE613989S, BWE612753S, BWE612758S, BWE612788S, BWE612872S, BWE612875S, BWE612891S, BWE612968S, BWE612992S, BWE613004S, BWE613009S, BWE613011S, BWE613028S, BWE613032S, BWE613038S, BWE613043S, BWE613049S, BWE613079S, BWE613108S, BWE613394S, BWE613402S, BWE613411S, BWE613415S, BWE613416S, BWE613418S, BWE613423S, BWE613440S, BWE613456S, BWE613499S, BWE613500S, BWE613501S, BWE613512S, BWE613540S, BWE613559S, BWE613567S, BWE613568S, BWE613569S, BWE613570S, BWE613571S, BWE613575S, BWE613619S, BWE613622S, BWE613623S, BWE613639S, BWE613641S, BWE613689S, BWE613692S, BWE613695S, BWE613708S, BWE613709S, BWE613710S, BWE613719S, BWE613723S, BWE613764S, BWE613789S, BWE613790S, BWE613791S, BWE613800S, BWE613802S, BWE613803S, BWE613822S, BWE613823S, BWE613834S, BWE613840S, BWE613842S, BWE613849S, BWE613908S, BWE613909S, BWE613911S, BWE613913S, BWE613918S, BWE613986S, BWE613994S, BWE613995S, BWE613348S, BWE613405S, BWE613408S, BWE613424S, BWE613760S, BWE613835S, BWE613852S, BWE613853S, BWE613914S, BWE613991S, BWE612903S, BWE612904S, BWE613015S, BWE613027S, BWE613399S, BWE613438S, BWE613491S, BWE613508S, BWE613509S, BWE613514S, BWE613546S, BWE613576S, BWE613577S, BWE613627S, BWE613646S, BWE613701S, BWE613793S, BWE613796S, BWE613798S, BWE613825S, BWE613828S, BWE613832S, BWE613833S, BWE613848S, BWE613851S, BWE613917S, BWE613720S, BWE613582S, BWE613583S, BWE613584S, BWE612740S, BWE612752S, BWE612777S, BWE612871S, BWE612878S, BWE612907S, BWE612942S, BWE613030S, BWE613031S, BWE613075S, BWE613284S, BWE613303S, BWE613386S, BWE613407S, BWE613422S, BWE613428S, BWE613510S, BWE613829S, BWE613830S, BWE613916S, BWE612925S, BWE612934S, BWE612937S, BWE612941S, BWE613046S, BWE613047S, BWE613048S, BWE613516S, BWE613517S, BWE613518S, BWE613714S, BWE613715S, BWE613766S, BWE613770S, BWE613808S, BWE613824S, BWE613920S, BWE613403S, BWE613792S, BWE612890S, BWE612995S, BWE612999S, BWE613054S, BWE613056S, BWE613395S, BWE613515S, BWE613548S, BWE613573S, BWE613700S, BWE613795S, BWE613826S, BWE613905S, BWE613912S, BWE613915S, BWE612744S, BWE613037S, BWE613396S, BWE613406S, BWE613417S, BWE613809S, BWE613333S, BWE613282S, BWE613727S, BWE613293S, BWE613340S, BWE613383S, BWE612860S, BWE612869S, BWE613134S, BWE613318S, BWE613322S, BWE613593S, BWE612706S, BWE613359S, BWE613643S, BWE613651S, BWE613722S, BWE613268S, BWE613347S, BWE613363S, BWE613601S, BWE613625S, BWE613763S, BWE612680S, BWE612828S, BWE613600S, BWE613637S, BWE613640S, BWE613266S, BWE612700S, BWE612702S, BWE613264S, BWE613694S, BWE613726S, BWE613642S, BWE612736S, BWE612755S, BWE612846S, BWE613270S, BWE613291S, BWE613331S, BWE613388S, BWE613325S, BWE612774S, BWE613309S, BWE613337S, BWE613349S, BWE613358S, BWE613771S, BWE612655S, BWE613377S, BWE613691S, BWE613385S, BWE612676S, BWE612845S, BWE613605S, BWE613606S, BWE613697S, BWE613277S, BWE612656S, BWE613381S, BWE613200S, BWE613294S, BWE613336S, BWE613338S, BWE613369S, BWE613379S, BWE612756S, BWE613768S, BWE613401S, BWE613410S, BWE612781S, BWE613334S, BWE613380S, BWE613286S, BWE613373S, BWE612762S, BWE612759S, BWE613611S, BWE613698S, BWE613718S, BWE613769S, BWE612648S, BWE612674S, BWE612699S, BWE612735S, BWE613133S, BWE613330S, BWE613339S, BWE613346S, BWE613360S, BWE613361S, BWE613362S, BWE613374S, BWE613586S, BWE613594S, BWE613595S, BWE613602S, BWE613647S, BWE613653S, BWE613657S, BWE613772S, BWE612671S, BWE612768S, BWE613263S, BWE613271S, BWE613281S, BWE613703S, BWE613356S, BWE613109S, BWE612710S, BWE612780S, BWE612805S, BWE612809S, BWE613614S, BWE613265S, BWE613274S, BWE612820S, BWE613419S, BWE613693S, BWE613696S, BWE613717S, BWE613649S, BWE613328S, BWE613335S, BWE613603S, BWE613660S, BWE612757S, BWE612787S, BWE613267S, BWE613375S, BWE613371S, BWE613762S, BWE612830S, BWE613314S, BWE613355S, BWE613367S, BWE613372S, BWE613587S, BWE613599S, BWE613661S, BWE613690S, BWE613706S, BWE612771S, BWE612834S, BWE613604S, BWE613332S, BWE612767S, BWE613273S, BWE613592S, BWE613272S, BWE613366S, BWE613607S, BWE613628S, BWE613650S, BWE613652S, BWE613699S, BWE613712S, BWE613716S, BWE612664S, BWE612669S, BWE613357S, BWE613609S, BWE613658S, BWE613767S, BWE613617S, BWE613280S, BWE612773S, BWE613376S, BWE613387S, BWE612819S, BWE613276S, BWE613414S, BWE612733S, BWE613193S, BWE613390S, BWE613919S, BWE613988S, BWE613292S, BWE613177S, BWE612725S, BWE612726S, BWE612727S, BWE612728S, BWE612729S, BWE612707S, BWE613590S, BWE613353S, BWE613612S, BWE613585S, BWE613850S, BWE613987S, BWE613608S, BWE612841S, BWE613261S, BWE613350S, BWE613618S, BWE612694S, BWE613737S, BWE612678S, BWE612863S, BWE613845S, BWE612711S, BWE613351S, BWE612691S, BWE613342S, BWE613596S, BWE612654S;  GTIN 00643169980204, Lot Serial Numbers:  BWE614420S, BWE614714S, BWE615090S, BWE615173S, BWE615194S, BWE613896S, BWE613897S, BWE613898S, BWE614007S, BWE614008S, BWE614025S, BWE614263S, BWE614351S, BWE614362S, BWE614363S, BWE614364S, BWE614399S, BWE614424S, BWE614552S, BWE614629S, BWE614931S, BWE615001S, BWE615022S, BWE615023S, BWE615177S, BWE615188S, BWE614404S, BWE614538S, BWE615020S, BWE614300S, BWE613684S, BWE614341S, BWE613741S, BWE613780S, BWE614278S, BWE614375S, BWE614548S, BWE613534S, BWE613779S, BWE613781S, BWE613954S, BWE613526S, BWE613674S, BWE613678S, BWE613751S, BWE613783S, BWE614037S, BWE614949S, BWE613729S, BWE614438S, BWE614383S, BWE614384S, BWE614436S, BWE614437S, BWE613536S, BWE613734S, BWE613782S, BWE615019S, BWE613754S, BWE614160S, BWE613676S, BWE614962S, BWE614963S, BWE614272S, BWE614317S, BWE614311S, BWE614596S, BWE614790S, BWE614849S, BWE614851S, BWE614853S, BWE614930S, BWE614961S, BWE614997S, BWE614998S, BWE614999S, BWE615033S, BWE615052S, BWE614919S, BWE614993S, BWE614992S, BWE614995S, BWE614996S, BWE615003S, BWE615028S, BWE615036S, BWE615039S, BWE615056S, BWE613136S, BWE613520S, BWE613531S, BWE613537S, BWE613733S, BWE614553S, BWE613527S, BWE613681S, BWE614177S, BWE614238S, BWE614299S, BWE614407S, BWE614427S, BWE614654S, BWE614754S, BWE614921S, BWE615035S, BWE615030S, BWE613671S, BWE614943S, BWE613135S, BWE613679S, BWE613730S, BWE613978S, BWE613979S, BWE614035S, BWE614250S, BWE614270S, BWE613670S, BWE614028S, BWE614315S, BWE613528S, BWE613673S, BWE614021S, BWE614026S, BWE614298S, BWE614342S, BWE614338S, BWE613523S, BWE613785S, BWE614354S, BWE614385S, BWE614423S, BWE614532S, BWE614627S, BWE614628S, BWE614789S, BWE614852S, BWE614865S, BWE614914S, BWE614922S, BWE614952S, BWE613524S, BWE613525S, BWE613551S, BWE613675S, BWE613538S, BWE613550S, BWE613552S, BWE613554S, BWE613669S, BWE613744S, BWE613746S, BWE613889S, BWE613890S, BWE613899S, BWE613900S, BWE613901S, BWE613930S, BWE613939S, BWE613940S, BWE613731S, BWE614027S, BWE613682S, BWE613519S, BWE613529S, BWE613530S, BWE613535S, BWE613553S, BWE613557S, BWE613672S, BWE613688S, BWE613872S, BWE613893S, BWE613895S, BWE613955S, BWE613956S, BWE614019S, BWE614030S, BWE614034S, BWE614086S, BWE614087S, BWE614158S, BWE614159S, BWE614280S, BWE614281S, BWE614291S, BWE614292S, BWE614314S, BWE614343S, BWE614346S, BWE614359S, BWE614360S, BWE614441S, BWE614463S, BWE614541S, BWE614542S, BWE614547S, BWE614554S, BWE614578S, BWE614611S, BWE614683S, BWE614699S, BWE614700S, BWE614748S, BWE614893S, BWE614709S, BWE614537S, BWE614074S, BWE614075S, BWE614080S, BWE614081S, BWE614082S, BWE614102S, BWE614103S, BWE614104S, BWE614105S, BWE614107S, BWE614109S, BWE614110S, BWE614111S, BWE614112S, BWE614129S, BWE614131S, BWE614135S, BWE614149S, BWE614151S, BWE614154S, BWE614194S, BWE614199S, BWE614204S, BWE614235S, BWE615195S, BWE614056S, BWE614069S, BWE614100S, BWE614226S, BWE614230S, BWE614295S, BWE614329S, BWE614334S, BWE614350S, BWE614585S, BWE614587S, BWE614641S, BWE614643S, BWE614705S, BWE614726S, BWE614907S, BWE614908S, BWE614935S, BWE614964S, BWE614965S, BWE615084S, BWE615096S, BWE615193S, BWE614120S, BWE614206S, BWE614113S, BWE614117S, BWE614052S, BWE614066S, BWE614076S, BWE614077S, BWE614089S, BWE614096S, BWE614097S, BWE614098S, BWE614114S, BWE614115S, BWE614123S, BWE614124S, BWE614125S, BWE614126S, BWE614198S, BWE614221S, BWE614243S, BWE614271S, BWE614303S, BWE614309S, BWE614347S, BWE614462S, BWE614519S, BWE614588S, BWE614606S, BWE614635S, BWE614648S, BWE614741S, BWE614905S, BWE614971S, BWE614980S, BWE614065S, BWE614059S, BWE614061S, BWE614068S, BWE614072S, BWE614084S, BWE614095S, BWE614101S, BWE614106S, BWE614157S, BWE614208S, BWE614209S, BWE614236S, BWE614239S, BWE614240S, BWE614245S, BWE614247S, BWE614249S, BWE614252S, BWE614266S, BWE614307S, BWE614318S, BWE614319S, BWE614321S, BWE614326S, BWE614336S, BWE614476S, BWE614493S, BWE614498S, BWE614520S, BWE614522S, BWE614577S, BWE614584S, BWE614586S, BWE614605S, BWE614634S, BWE614636S, BWE614645S, BWE614755S, BWE614803S, BWE614842S, BWE614843S, BWE614909S, BWE614940S, BWE614953S, BWE614958S, BWE615182S, BWE615184S, BWE615190S, BWE614057S, BWE614064S, BWE614070S, BWE614071S, BWE614090S, BWE614153S, BWE614155S, BWE614161S, BWE614213S, BWE614216S, BWE614242S, BWE614260S, BWE614265S, BWE614279S, BWE614286S, BWE614310S, BWE614322S, BWE614328S, BWE614331S, BWE614472S, BWE614640S, BWE614647S, BWE614093S, BWE614148S, BWE614150S, BWE614264S, BWE614058S, BWE614108S, BWE614067S, BWE614073S, BWE614078S, BWE614079S, BWE614116S, BWE614127S, BWE614130S, BWE614146S, BWE614152S, BWE614156S, BWE614220S, BWE614244S, BWE614258S, BWE614261S, BWE614268S, BWE614269S, BWE614304S, BWE614305S, BWE614306S, BWE614323S, BWE614332S, BWE614345S, BWE614349S, BWE614515S, BWE614523S, BWE614598S, BWE614621S, BWE614633S, BWE614637S, BWE614812S, BWE614838S, BWE614902S, BWE614956S, BWE614960S, BWE614665S, BWE614666S, BWE614667S, BWE614668S, BWE614669S, BWE614694S, BWE614702S, BWE614722S, BWE614725S, BWE614727S, BWE614729S, BWE614733S, BWE614739S, BWE614744S, BWE614756S, BWE614757S, BWE614758S, BWE614765S, BWE614766S, BWE614771S, BWE614774S, BWE614776S, BWE614796S, BWE614801S, BWE614802S, BWE614804S, BWE614806S, BWE614809S, BWE614810S, BWE614837S, BWE614841S, BWE614844S, BWE614807S, BWE614494S, BWE614543S, BWE614659S, BWE614686S, BWE614687S, BWE614689S, BWE614704S, BWE614708S, BWE614738S, BWE614745S, BWE614760S, BWE614767S, BWE614768S, BWE614770S, BWE614826S, BWE614827S, BWE614836S, BWE614710S, BWE614544S, BWE614545S, BWE614546S, BWE614549S, BWE614569S, BWE614574S, BWE614576S, BWE614580S, BWE614663S, BWE614664S, BWE614685S, BWE614688S, BWE614713S, BWE614716S, BWE614717S, BWE614718S, BWE614723S, BWE614724S, BWE614798S, BWE614800S, BWE614513S, BWE614720S, BWE614675S, BWE614255S, BWE614624S, BWE614555S, BWE614559S, BWE614561S, BWE614879S, BWE613959S, BWE614184S, BWE614377S, BWE614386S, BWE614425S, BWE614083S, BWE613756S, BWE613778S, BWE613868S, BWE613963S, BWE613974S, BWE613982S, BWE614009S, BWE614171S, BWE614172S, BWE614267S, BWE614366S, BWE614379S, BWE614380S, BWE614416S, BWE614524S, BWE614525S, BWE614408S, BWE613980S, BWE614192S, BWE614237S, BWE614557S, BWE614568S, BWE614599S, BWE614984S, BWE613860S, BWE613861S, BWE613967S, BWE614032S, BWE614169S, BWE614372S, BWE614410S, BWE614426S, BWE614521S, BWE614570S, BWE614615S, BWE614616S, BWE614617S, BWE614631S, BWE614889S, BWE614910S, BWE613752S, BWE613929S, BWE614038S, BWE614168S, BWE614417S, BWE614419S, BWE613925S, BWE613976S, BWE614012S, BWE614022S, BWE614023S, BWE614033S, BWE614137S, BWE614186S, BWE614376S, BWE614447S, BWE614453S, BWE614556S, BWE614564S, BWE615025S, BWE614128S, BWE614133S, BWE614449S, BWE614421S, BWE613960S, BWE614387S, BWE614288S, BWE614289S, BWE614382S, BWE614626S, BWE614890S, BWE614015S, BWE614275S, BWE614138S, BWE614241S, BWE614320S, BWE614324S, BWE614378S, BWE614560S, BWE613864S, BWE613866S, BWE613964S, BWE613965S, BWE613973S, BWE613975S, BWE613981S, BWE614010S, BWE614011S, BWE614085S, BWE614136S, BWE614166S, BWE614185S, BWE614190S, BWE614233S, BWE614330S, BWE614356S, BWE614409S, BWE614428S, BWE614445S, BWE614529S, BWE614530S, BWE614620S, BWE614622S, BWE614563S, BWE615130S, BWE614753S, BWE613952S, BWE614517S, BWE614039S, BWE614132S, BWE614134S, BWE614676S, BWE614031S, BWE614863S, BWE614892S, BWE614029S, BWE614174S, BWE614191S, BWE614355S, BWE614448S, BWE614455S, BWE614518S, BWE614013S, BWE614024S, BWE614365S, BWE614369S, BWE613871S, BWE614063S, BWE614325S, BWE614327S, BWE614368S, BWE614446S, BWE614337S, BWE614251S, BWE614358S, BWE614189S, BWE615088S, BWE615031S, BWE613894S, BWE614461S, BWE614591S, BWE613855S, BWE614273S, BWE615026S, BWE614187S, BWE614381S, BWE614418S, BWE614878S, BWE613873S, BWE613891S, BWE613902S, BWE614277S, BWE614282S, BWE614371S, BWE614862S, BWE615047S, BWE613977S, BWE613757S, BWE614139S, BWE614367S, BWE614527S, BWE614562S, BWE614916S, BWE615024S, BWE613858S, BWE614118S, BWE614143S, BWE614167S, BWE614316S, BWE614333S, BWE614340S, BWE614388S, BWE614422S, BWE614526S, BWE614528S, BWE614531S, BWE614877S, BWE615050S, BWE613962S, BWE613970S, BWE614040S, BWE614373S, BWE614536S, BWE614565S, BWE614597S, BWE614613S, BWE614274S, BWE614823S, BWE614734S, BWE614736S, BWE614797S, BWE614824S, BWE614825S, BWE614845S, BWE614854S, BWE614121S, BWE614122S, BWE613747S, BWE614339S, BWE614751S, BWE614353S, BWE614614S, BWE615004S, BWE615029S, BWE615032S, BWE615034S, BWE615037S, BWE615038S, BWE614558S, BWE614701S, BWE613533S, BWE614374S, BWE614566S, BWE614721S, BWE613787S, BWE614711S, BWE614915S, BWE614036S, BWE614178S, BWE614698S, BWE614749S, BWE614750S, BWE614232S, BWE615102S, BWE615103S, BWE615104S, BWE615105S, BWE615106S, BWE615107S, BWE615108S, BWE615109S, BWE615110S, BWE615111S, BWE615112S, BWE615113S, BWE615114S, BWE615115S, BWE615116S, BWE615120S, BWE615121S, BWE615122S, BWE615123S, BWE615124S, BWE615125S, BWE615126S, BWE615127S, BWE615128S, BWE615129S, BWE614658S, BWE614660S, BWE614661S, BWE614016S, BWE614393S, BWE614433S, BWE614451S, BWE614452S, BWE614454S, BWE614477S, BWE614478S, BWE614638S, BWE614639S, BWE614655S, BWE614656S, BWE614657S, BWE614674S, BWE614695S, BWE614696S, BWE613686S, BWE613934S, BWE613936S, BWE613938S, BWE613942S, BWE613943S, BWE613946S, BWE613947S, BWE613948S, BWE613951S, BWE614047S, BWE614048S, BWE614049S, BWE614050S, BWE614592S, BWE614593S, BWE614594S, BWE614595S, BWE614920S, BWE614170S, BWE615002S, BWE614539S, BWE613728S, BWE614435S, BWE614450S, BWE614731S, BWE614632S, BWE614679S, BWE614684S, BWE614690S, BWE614691S, BWE614692S, BWE614707S, BWE614728S, BWE614737S, BWE614764S, BWE614821S, BWE614822S, BWE614828S, BWE614829S, BWE614583S, BWE614099S, BWE614091S, BWE613928S, BWE613859S, BWE613854S, BWE613931S, BWE613958S, BWE614608S, BWE614088S, BWE614491S, BWE614983S, BWE614985S, BWE614986S, BWE615077S, BWE613961S, BWE614858S, BWE614870S, BWE614195S, BWE614196S, BWE614197S, BWE614200S, BWE614201S, BWE614202S, BWE614203S, BWE614205S, BWE614207S, BWE614210S, BWE613966S, BWE614051S, BWE614053S, BWE614054S, BWE614055S, BWE614060S, BWE614062S, BWE614140S, BWE614141S, BWE613862S, BWE614370S, BWE614165S, BWE614179S, BWE614180S, BWE614181S, BWE614182S, BWE614290S, BWE614293S, BWE614294S, BWE614297S, BWE614335S, BWE614389S, BWE614390S, BWE614391S, BWE614392S, BWE614395S, BWE614396S, BWE614442S, BWE614443S, BWE614444S, BWE614456S, BWE614503S, BWE614504S, BWE614505S, BWE614506S, BWE614540S, BWE614603S, BWE614604S, BWE614607S, BWE614680S, BWE614682S, BWE614697S, BWE614703S, BWE614785S, BWE614788S, BWE614791S, BWE614792S, BWE614805S, BWE614813S, BWE614814S, BWE614815S, BWE614816S, BWE614817S, BWE614855S, BWE614856S, BWE614868S, BWE614869S, BWE614873S, BWE614874S, BWE614875S, BWE614876S, BWE614880S, BWE614881S, BWE614882S, BWE614883S, BWE614896S, BWE614901S, BWE614903S, BWE614904S, BWE614906S, BWE614911S, BWE614978S, BWE614979S, BWE614981S, BWE614982S, BWE614990S, BWE615005S, BWE615006S, BWE615012S, BWE615013S, BWE615040S, BWE615041S, BWE615042S, BWE615043S, BWE615044S, BWE615045S, BWE615046S, BWE615057S, BWE615065S, BWE615066S, BWE615067S, BWE615068S, BWE615069S, BWE615070S, BWE615092S, BWE615093S, BWE615094S, BWE615095S, BWE615097S, BWE615098S, BWE615099S, BWE615145S, BWE615146S, BWE615147S, BWE615148S, BWE615149S, BWE615150S, BWE615151S, BWE615152S, BWE615154S, BWE615155S, BWE614831S, BWE614832S, BWE614778S, BWE614793S, BWE614820S, BWE615089S, BWE615100S, BWE615101S, BWE614735S, BWE614775S, BWE614808S, BWE614811S, BWE613983S, BWE614740S, BWE614550S, BWE613735S, BWE613739S, BWE613784S, BWE613863S, BWE613867S, BWE613870S, BWE613874S, BWE613877S, BWE614730S, BWE614211S, BWE614094S, BWE614840S, BWE614929S, BWE614693S, BWE614497S, BWE614501S, BWE613555S, BWE614928S, BWE614500S, BWE614018S, BWE613556S, BWE614839S, BWE615000S, BWE614014S, BWE615082S, BWE613558S, BWE613880S, BWE613885S, BWE613886S, BWE613957S, BWE613685S, BWE613968S, BWE613969S, BWE613971S, BWE613972S, BWE614041S, BWE614042S, BWE614043S, BWE614044S, BWE614045S, BWE614046S, BWE614144S, BWE614145S, BWE614147S, BWE614163S, BWE614164S, BWE614175S, BWE614176S, BWE614183S, BWE614256S, BWE614257S, BWE614259S, BWE614262S, BWE614361S, BWE614397S, BWE614398S, BWE614400S, BWE614401S, BWE614402S, BWE614403S, BWE614411S, BWE614412S, BWE614413S, BWE614414S, BWE614415S, BWE614429S, BWE614430S, BWE614431S, BWE614432S, BWE614434S, BWE614439S, BWE614440S, BWE614457S, BWE614458S, BWE614459S, BWE614460S, BWE614464S, BWE614465S, BWE614466S, BWE614467S, BWE614468S, BWE614469S, BWE614470S, BWE614471S, BWE614473S, BWE614475S, BWE614479S, BWE614480S, BWE614481S, BWE614482S, BWE614483S, BWE614484S, BWE614485S, BWE614486S, BWE614487S, BWE614488S, BWE614489S, BWE614490S, BWE614495S, BWE614496S, BWE614502S, BWE614511S, BWE614512S, BWE614514S, BWE614600S, BWE614601S, BWE614602S, BWE614779S, BWE614781S, BWE614782S, BWE614783S, BWE614784S, BWE614818S, BWE614819S, BWE614871S, BWE614872S, BWE614884S, BWE614887S, BWE614888S, BWE614894S, BWE614895S, BWE614923S, BWE614924S, BWE614925S, BWE614926S, BWE614927S, BWE614932S, BWE614933S, BWE614934S, BWE614936S, BWE614938S, BWE614939S, BWE614941S, BWE614942S, BWE614944S, BWE614945S, BWE614946S, BWE614947S, BWE614950S, BWE614955S, BWE614957S, BWE614966S, BWE614968S, BWE614969S, BWE614972S, BWE614973S, BWE614974S, BWE614975S, BWE614987S, BWE614988S, BWE614989S, BWE615071S, BWE615072S, BWE615073S, BWE615074S, BWE615075S, BWE615076S, BWE614092S, BWE614662S, BWE614579S, BWE613786S, BWE614715S, BWE614794S, BWE615083S, BWE614229S, BWE613941S, BWE614006S, BWE614017S, BWE614848S, BWE615086S, BWE614405S, BWE614406S, BWE614499S, BWE614551S, BWE614976S, BWE614533S, BWE614534S, BWE614847S, BWE614712S, BWE614759S, BWE614763S, BWE614769S, BWE614773S, BWE614795S, BWE614833S, BWE614834S, BWE614835S, BWE615017S, BWE615018S, BWE615141S, BWE614507S, BWE614508S, BWE614509S, BWE614510S, BWE614516S, BWE614743S, BWE614746S, BWE614857S, BWE614859S, BWE614860S, BWE614861S, BWE614864S, BWE614866S, BWE614867S, BWE615007S, BWE615008S, BWE615009S, BWE615010S, BWE615011S, BWE615014S, BWE615015S, BWE615016S, BWE615078S, BWE615079S, BWE615080S, BWE615081S, BWE614762S, BWE614885S, BWE614897S, BWE614898S, BWE614899S, BWE614948S, BWE614670S, BWE614671S, BWE614672S, BWE614673S, BWE614677S, BWE614678S, BWE614719S, BWE614772S, BWE614777S, BWE614799S, BWE614830S, BWE614846S, BWE614732S, BWE615085S, BWE615087S, BWE615091S, BWE615170S, BWE615172S, BWE615176S, BWE615181S, BWE615192S, BWE615197S",
      "more_code_info": ""
    }
  ]
}