  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76656",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including the countries of:   Australia  Belgium  Canada  Denmark  Finland  France  Germany  Germany  India  Italy  Netherlands  Norway  Qatar  Singapore  South Africa  Sweden  Switzerland  United Kingdom",
      "recall_number": "Z-1707-2017",
      "product_description": "Philips IntelliVue MX4O WLAN Patient Wearable Monitor  Product: 865352  Exchange part (service numbers):  453564615311 TELE PWM,802.lla/b/g,ECG only, US only  453564615331 TELE PWM,802.1 1 a/b/g,ECG&Sp02, US only    Product Usage:  Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX4O is connected the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX4O does not automatically provide local monitoring or alarming when connected to the Information Center.",
      "product_quantity": "2648",
      "reason_for_recall": "Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to  Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]",
      "recall_initiation_date": "20170306",
      "center_classification_date": "20170404",
      "termination_date": "20180525",
      "report_date": "20170412",
      "code_info": "The following product number and Exchange part numbers containing the combination of HW Revision C.01 .01 with SW Revisions B.05 or B.06 of the Philips IntelliVue MX4O Patient Wearable Monitor"
    }
  ]
}