  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71054",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Drive",
      "address_2": "N/A",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Class I Recall - Worldwide distribution -- US, including the states of AL, AZ, DC, FL, MI, MO, NJ, NY, OR, PA, TX, and WV; and, countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Libya, Mexico, Mongolia, Mozambique, Netherlands, Norway Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Unites States, and Venezuela.",
      "recall_number": "Z-1706-2015",
      "product_description": "FLOW-i Anesthesia System    The FLOW-i Anesthesia System is intended for use administering inhalation anesthesia on neonatal to adult patient populations.",
      "product_quantity": "1641 [80 units (US) 1561 units (OUS)]",
      "reason_for_recall": "In some Flow-i Anesthesia Systems, patient cassette can get dislodged which could result in gas leakage within the system internal circuit potentially leading to a stop in ventilation.",
      "recall_initiation_date": "20150402",
      "center_classification_date": "20150608",
      "termination_date": "20160706",
      "report_date": "20150617",
      "code_info": "Article number  66 77 200  66 77 300  66 77 400    Serial Numbers  1058 1646 2034 2723  1059 1647 2095 2724  1060 1648 2186 2729  1172 1650 2554 2731  1201 1651 2555 2732  1202 1652 2556 2733  1204 1653 2557 2735  1224 1654 2558 2736  1225 1655 2559 2737  1226 1656 2560 2738  1277 1657 2561 2739  1280 1786 2562 2740  1281 1787 2613 2741  1282 1860 2714 2742  1545 1861 2716 2743  1546 2021 2719 2744  1643 2022 2720 2745  1644 2023 2721 2746  1645 2024 2722 2747"
    }
  ]
}