      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79987",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide Distribution - US Distribution to the states of : CA, FL, MO, NC, SC, TX, and WI., and to the countries of :  Argentina, China, Colombia, Germany, Japan, Netherlands, South Africa, Thailand, and Vietnam.",
      "recall_number": "Z-1705-2018",
      "product_description": "Lactosorb RapidFlap, bone plate, Model No. 915-0020    The RapidFlap LS Cranial Fixation System is indicated for use in pediatric craniotomy flap fixation.",
      "product_quantity": "13175",
      "reason_for_recall": "The recalling firm has confirmed that the Outer Plate component exhibits an excessive chamfer on the threading after deburring operations. This excessive chamfer results in non-conforming product where the threads of the outer plate component have limited to no engagement with the post component.",
      "recall_initiation_date": "20180405",
      "center_classification_date": "20180509",
      "termination_date": "20200812",
      "report_date": "20180516",
      "code_info": "Lots 013040  013050  013060  013070  013080  013090  013100  013120  013130  013140  013150  025720  038090  038100  038110  038120  038130  038160  038170  038200  038210  038220  038230  054050  054070  054080  054090  054110  075970  089200  089210  108270  108280  110830  110840  110860  110870  110900  110910  110920  110930  110940  110950  110980  110990  111000  111010  111020  111030  111070  144270  144280  147300  147350  165500  165510  173190  173210  173220  173230  173260  173270  173280  173290  173300  173320  173340  218330  218340  255340  255370  264730  264750  264760  264770  264780  291390  291400  293380  293390  293400  293410  293470  293480  315430  315450  315460  315470  315500  315520  315530  315540  315560  331750  331800  331820  331840  331850  331880  335580  335590  335600  335610  348850  348860  348870  379760  379800  379810  379870  379920  389450  389460  389470  404640  404650  404660  443260  443270  443280  444360  444380  444390  444400  444420  456140  456340  456350  456370  467210  467220  469270  469280  469300  469310  471700  471720  471760  479790  479810  479820  479840  480560  480620  480660  508530  508540  525820  525830  525850  525860  543340  543370  543380  543390  543400  550440  550450  550460  550470  559890  559900  559910  559920  559930  559950  559960  559980  559990  562220  562230  562240  562260  563380  563390  563400  581190  581200  669150  669160  669180  669190  669200  675730  676640  676650  676720  676730  676740  676750  676760  676780  676790  676800  676820  676830  676840  676850  691060  691100  691110  691230  691240  691250  691290  691300  691320  701210  701220  701230  701240  701250  701270  701280  715290  715300  715310  715320  715330  715340  788400  788410  788420  788430  788440  792970  792990  793000  793010  793050  793080  793090  793100  810070  810100  810110  835080  835090  835100  835110  835120  835140  875960  876060  876070  876090  876100  876110  876120  876130  876150  898160  898170  898200  898240  898260  898320  898360  914430  914450  914460  930760  930780  930810  930840  930850  930860  930870  946980  947030  947070  991710  991740  991760  991770  991780  997070  997110  997120  997130  997150  997160  997170  997190  997260  997300  997320  997340  997360  997390  997400  997410    (Updated 3/20/19)  Expanded Lots:  538220  593400  331850  538510  593410  379800  538880  593430  691290  538900  615370  054070  538940  615380  110860  662300  615390  110900  699490  615400  110940  699600  615410  111030  991750  615420  173290  379730  615630  218330  379910  662320  218340  538120  662330  331750  538180  662370  331820  538250  662380  331840  538310  699470  331880  538370  699500  379760  538480  699520  379810  538520  699540  379870  538610  699560  379920  538630  699570  444380  538660  699580  676830  538680  699590  835110  538920  110910  997260"
    }
  ]
}