  "results": [
    {
      "country": "United States",
      "city": "Snoqualmie",
      "address_1": "35301 SE Center St",
      "reason_for_recall": "A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.",
      "address_2": "",
      "product_quantity": "64 devices",
      "code_info": "All devices of Model: 91496.",
      "center_classification_date": "20140603",
      "distribution_pattern": "US Distribution to GA.",
      "state": "WA",
      "product_description": "Spacelabs Medical Ultraview SL Command Module, Model 91496.     Arrhythmia Detector and Alarm. The  Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Spacelabs Healthcare Inc",
      "recall_number": "Z-1705-2014",
      "initial_firm_notification": "Visit",
      "product_type": "Devices",
      "event_id": "68357",
      "termination_date": "20140603",
      "more_code_info": "",
      "recall_initiation_date": "20140508",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",