  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Buffalo Grove",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73823",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "1700 Leider Ln",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- U.S., to the following states: AK, CA, FL, GA, MN, MT, NJ, OH, TX and WA; and, the country of Canada.",
      "recall_number": "Z-1704-2016",
      "product_description": "Novocastra Liquid Mouse Monoclonal Antibody Calretinin, NCL-L-CALRET-566.    Intended for the qualitative identification by light microscopy of human calretinin molecule in paraffin sections.",
      "product_quantity": "290 units",
      "reason_for_recall": "Novocastra Liquid Mouse Monoclonal Antibody Calretinin when used in combination with specific Lots of antibody diluent will give reduced staining.",
      "recall_initiation_date": "20160322",
      "center_classification_date": "20160520",
      "termination_date": "20171130",
      "report_date": "20160601",
      "code_info": "Lot numbers: 6025204, 6029060, 6029949, 6035923 and 6036060."
    }
  ]
}