  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lincolnshire",
      "address_1": "577 Aptakisic Rd",
      "reason_for_recall": "Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.",
      "address_2": "",
      "product_quantity": "93 units.",
      "code_info": "Catalog No. AH699752.",
      "center_classification_date": "20200414",
      "distribution_pattern": "USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..",
      "state": "IL",
      "product_description": "The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.",
      "report_date": "20200422",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Sysmex America, Inc.",
      "recall_number": "Z-1703-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85254",
      "termination_date": "20210324",
      "more_code_info": "",
      "recall_initiation_date": "20200309",
      "postal_code": "60069-4325",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}