  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Recovery for assayed controls are out of assigned ranges.  A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed.  Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.",
      "address_2": "PO BOX 6101",
      "product_quantity": "27501",
      "code_info": "Catalog Number B4218-1, Material Number 10445709, lot numbers 557128, 557129, 557130,  557131, 557132, 557133,  557134, 557135, 557140,  557141 , expiration dates:10/31/2015, 11/21/2015, 2/16/2016",
      "center_classification_date": "20150602",
      "distribution_pattern": "Worldwide Distribution.  US naionwide including Puerto Rico, Barbados and Mexico.",
      "state": "DE",
      "product_description": "Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.",
      "report_date": "20150610",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-1703-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71273",
      "termination_date": "20170124",
      "recall_initiation_date": "20150325",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}