  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carol Stream",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79972",
      "recalling_firm": "Bard Brachytherapy, Inc.",
      "address_1": "295 E Lies Rd",
      "address_2": "N/A",
      "postal_code": "60188-9421",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Distribution to the states of : California and Georgia",
      "recall_number": "Z-1702-2018",
      "product_description": "BARD¿ BrachySource¿ Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 absorbed onto a nickel / copper-coated, gold-cored aluminum wire (reference Figure 1 below). The I125 has a half-life of 59.6 days and decays by electron capture with the emission of characteristic photons and Auger electrons. The principal photon emissions are 27.4 and 31 keV x-rays and 35.5 keV gamma. The titanium wall of the BrachySource¿ Seed Implants absorbs the electrons.",
      "product_quantity": "70 Seeds at 0.28mCi and 100 seeds at 0.36mCi",
      "reason_for_recall": "Bard Medical Division is recalling the BARD¿ BrachySource¿ Iodine125 Radioactive Seeds in Mick¿ Applicators and BARD¿ BrachySource¿ Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource¿ Iodine125 0.36 mCi radioactive seeds but instead received BrachySource¿ Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.",
      "recall_initiation_date": "20180410",
      "center_classification_date": "20180507",
      "termination_date": "20200605",
      "report_date": "20180516",
      "code_info": "Product Code:1251CSR, BrachySource¿ Seed Implants in Re-Usable Mick¿ Applicators (60 seeds at 0.28mCi) with Lot BBCP0079    Product Code: 1251CSRA, BrachySource¿ Seed Implants in Shielded Mick¿ Applicators (10 seeds at 0.28mCi) with Lot BBCQ0026    Product Code: PS1251LS, Loose BrachySource¿ Seed Implants (with Lot BBCQ0050) Spacers (with Lot 23498276) 100 Seeds at 0.36 mCi."
    }
  ]
}