  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Danbury",
      "address_1": "36 Apple Ridge Rd # 37",
      "reason_for_recall": "Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury",
      "address_2": "",
      "product_quantity": "3 units",
      "code_info": "Serial Numbers: 1800314 7263, 180031472, 18002023051PR",
      "center_classification_date": "20140603",
      "distribution_pattern": "Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.",
      "state": "CT",
      "product_description": "Hologic M-IV Mammography System; Model M4--00001.    Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hologic, Inc.",
      "recall_number": "Z-1701-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68298",
      "termination_date": "20140818",
      "more_code_info": "",
      "recall_initiation_date": "20140513",
      "postal_code": "06810-7301",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}