  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85275",
      "recalling_firm": "Abbott",
      "address_1": "15900 Valley View Ct",
      "address_2": "",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV.    OUS:  None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.",
      "recall_number": "Z-1700-2020",
      "product_description": "Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical    The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.",
      "product_quantity": "Software Version 24.6.1 installed into 4733 programmers",
      "reason_for_recall": "Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.",
      "recall_initiation_date": "20200228",
      "center_classification_date": "20200410",
      "termination_date": "20240418",
      "report_date": "20200422",
      "code_info": "Software Version: 24.6.1    Model: 3650, GTIN: 05414734500661    Model: 3330, GTIN: 05414734509725    For affected serial number of programmers see Attachment B.1 Programmers    For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.",
      "more_code_info": ""
    }
  ]
}