  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73441",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Distribution.",
      "recall_number": "Z-1700-2016",
      "product_description": "1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack  (Unique Device Identifier No. 10758750002320, 20758750002327) and the 2)VITROS¿ Immunodiagnostic Products Total Beta-hCG II Calibrators (Unique Device Identifier No. 10758750002337, 20758750002334); For quantitative measurement of human chorionic gonadotropin (hCG) and its beta subunit in human serum and plasma.",
      "product_quantity": "14,780 pieces",
      "reason_for_recall": "When testing was conducted using certain lots, VITROS¿ Systems generated results within the measuring range for samples known to not contain measurable hCG. The system reported results up to approximately 7.40 mIU/mL (IU/L) for patient samples that should have been less than the measuring range of the assay (<2.39 mIU/mL (IU/L)).",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20160520",
      "termination_date": "20180530",
      "report_date": "20160601",
      "code_info": "Product Codes: 6802220, 6802221; Affected Lot Numbers: 1410, 1420, 1430 (With Expiry Date 17-Jun-2016), Lot Number: 1440 (With Expiry Date 15-Aug-2016, Lot Numbers: 1450, 1460, 1470 (With Expiry Date 17-Aug-2016) and Lot Number 1480 distributed only outside of U.S. (With Expiry Date 30-Sept-2016)"
    }
  ]
}