      "country": "United States",
      "city": "Danbury",
      "address_1": "36 Apple Ridge Rd # 37",
      "reason_for_recall": "Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury",
      "address_2": "",
      "product_quantity": "11 units",
      "code_info": "Serial Numbers:  2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067  2840314W8073, 2840314W8064, 2840314W8056W, 28401082720  28407083579, 28401137611W, 28403083045W",
      "center_classification_date": "20140603",
      "distribution_pattern": "Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.",
      "state": "CT",
      "product_description": "Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039.    The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hologic, Inc.",
      "recall_number": "Z-1700-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68298",
      "termination_date": "20140818",
      "more_code_info": "",
      "recall_initiation_date": "20140513",
      "postal_code": "06810-7301",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}