      "country": "United States",
      "city": "Buffalo Grove",
      "address_1": "1700 Leider Ln",
      "reason_for_recall": "Novocastra Liquid Mouse Monoclonal Antibody  Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.",
      "address_2": "",
      "product_quantity": "33 units",
      "code_info": "Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06",
      "center_classification_date": "20140602",
      "distribution_pattern": "Nationwide Distribution including AZ, CA, FL, and NY.",
      "state": "IL",
      "product_description": "Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant.   For in vitro diagnostic use.",
      "report_date": "20140611",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Leica Microsystems, Inc.",
      "recall_number": "Z-1698-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68065",
      "termination_date": "20170814",
      "more_code_info": "",
      "recall_initiation_date": "20140226",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}