  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94351",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution in the countries of Portugal and Italy.",
      "recall_number": "Z-1697-2024",
      "product_description": "Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation and/or ligation.",
      "product_quantity": "36 total, OUS distribution only",
      "reason_for_recall": "One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.",
      "recall_initiation_date": "20240125",
      "center_classification_date": "20240429",
      "report_date": "20240508",
      "code_info": "UDI-DI 20884521086422 Lot D2F2101RY",
      "more_code_info": ""
    }
  ]
}