  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Madison",
      "address_1": "1240 Deming Way",
      "reason_for_recall": "There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows.  Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.",
      "address_2": "",
      "product_quantity": "68 units",
      "code_info": "a) 0660000: CyberKnife Treatment Delivery System  Serial Numbers:  C0434, C0436, C0437, C0438, C0440, C0441, C0443, C0444, C0445, C0446, C0447, C0448, C0449, C0450, C0453, C0454, C0456, C0457, C0458, C0460, C0464, C0465, C0466, C0472, C0475      b) 053101-010:  CyberKnife M6 FI Treatment Delivery System  Serial Numbers: C0334, C0381    c) 053201-010:  CyberKnife M6 FM Treatment Delivery System  Serial Numbers:  C0363    d) 053301-010:  CyberKnife M6 FIM Treatment Delivery System  Serial Numbers:  C0337, C0343, C0349, C0352, C0358, C0370, C0371, C0378, C0384, C0387, C0389, C0390, C0391, C0393, C0409, C0424, C0426, C0427, C0428, C0433, C0439      e) 053401-010:  CyberKnife M6 FI+ Treatment Delivery System  Serial Numbers:  C0330, C0332, C0342, C0344, C0345, C0348, C0359, C0360, C0364, C0367, C0372, C0394, C0398, C0400, C0402, C0404, C0415, C0419      f) 053501-010:  CyberKnife Matrix Tier F TDS  Serial Numbers:  C0396",
      "center_classification_date": "20190531",
      "distribution_pattern": "US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand",
      "state": "WI",
      "product_description": "CyberKnife M6, Part Numbers:  a) 0660000: CyberKnife Treatment Delivery System  b) 053101-010:  CyberKnife M6 FI Treatment Delivery System  c) 053201-010:  CyberKnife M6 FM Treatment Delivery System  d) 053301-010:  CyberKnife M6 FIM Treatment Delivery System  e) 053401-010:  CyberKnife M6 FI+ Treatment Delivery System  f) 053501-010:  CyberKnife Matrix Tier 2 TDS",
      "report_date": "20190612",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Accuray Incorporated",
      "recall_number": "Z-1696-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82664",
      "termination_date": "20210401",
      "more_code_info": "",
      "recall_initiation_date": "20180410",
      "postal_code": "53717-1954",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}