  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71250",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including PR and the states of AL, AZ, CA, CO, FL, IL, IN, KS, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, and the countries of Canada, Korea, Singapore, Australia, Japan, Malaysia, India, Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Iceland, Iran, Israel, Italy, Norway, Russian Federation, Saudi Arabia, South America, Spain, Sweden, Switzerland, United Kingdom, and United Arab Emirates.",
      "recall_number": "Z-1696-2015",
      "product_description": "PRI Femoral Impactor Block for knee prosthesis surgery.    Orthopedic Manual Surgical Instrument",
      "product_quantity": "1,689",
      "reason_for_recall": "Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.",
      "recall_initiation_date": "20150430",
      "center_classification_date": "20150529",
      "termination_date": "20160315",
      "report_date": "20150610",
      "code_info": "Item No. 00-5901-032-00    Lots:  56572845  56574197  56574534  56574775  56575838  56573249  56574198  56574535  56575827  56575839  56573282  56574325  56574536  56575828  56575840  56573283  56574326  56574604  56575829  56576089  56573284  56574327  56574605  56575830  56576090  56573285  56574328  56574606  56575831  56576091  56574130  56574329  56574607  56575832  56576421  56574135  56574530  56574608  56575833  56576429  56574136  56574531  56574617  56575836  56576431  56574137  56574532  56574714  56575837  56576544  56574196  56574533  56574774"
    }
  ]
}