{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warwick",
      "state": "RI",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92276",
      "recalling_firm": "Mechatronics USA",
      "address_1": "20 Altieri Way Unit 4",
      "address_2": "",
      "postal_code": "02886-1761",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the state of OH and the country of  Netherlands.",
      "recall_number": "Z-1695-2023",
      "product_description": "RR Mechatronics  Control Level A  ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures.  Product Code: A0026578",
      "product_quantity": "10752 units",
      "reason_for_recall": "QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.",
      "recall_initiation_date": "20230414",
      "center_classification_date": "20230607",
      "report_date": "20230614",
      "code_info": "UDI:  (01)08719189137125(17)240222(10)QCAB6AAA28 Lot Number/Exp. Date: QCAB6AAA28 24-02-2022",
      "more_code_info": ""
    }
  ]
}