      "country": "United States",
      "city": "Medina",
      "address_1": "711 Park Ave",
      "reason_for_recall": "SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a \u001cSharp Watchdog Timeout\u001d error message.  Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pump\u0019s Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector.  For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.",
      "address_2": "",
      "product_quantity": "Domestic: 14,175 units; Canada: 187 units",
      "code_info": "14,362 Serial Numbers (please contact CDRH for list of affected Serial Numbers).",
      "center_classification_date": "20170329",
      "distribution_pattern": "Natiowide, Puerto Rico and Canada",
      "state": "NY",
      "product_description": "Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2",
      "report_date": "20170405",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Baxter Healthcare Corporation",
      "recall_number": "Z-1693-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76364",
      "termination_date": "20180524",
      "more_code_info": "",
      "recall_initiation_date": "20170126",
      "postal_code": "14103-1036",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}