      "country": "United States",
      "city": "Houston",
      "address_1": "14425 Torrey Chase Blvd Ste 100",
      "reason_for_recall": "The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image",
      "address_2": "",
      "product_quantity": "5 installed in US",
      "code_info": "NeuViz 16/64 Series",
      "center_classification_date": "20150616",
      "distribution_pattern": "Worldwide Distribution - US Distribution including Puerto Rico and to the states of : IL, MA, TX and SC., and to the countries of : China, Malaysia, Lebanon, Ecuador, Egypt, United Arab Emirates,Kazashstan, Tunisia, Angola, Brazil, Uzbekistan, Lebanon, Cote D' ivoire, South Africa, Montenegro and Tajikistan.",
      "state": "TX",
      "product_description": "Neusoft NeuViz 64 Multi-slice CT Scanner System",
      "report_date": "20150624",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Neusoft Medical System Co",
      "recall_number": "Z-1693-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71306",
      "termination_date": "20161205",
      "more_code_info": "",
      "recall_initiation_date": "20140308",
      "postal_code": "77014-1637",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}