  "results": [
    {
      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79505",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Canada, Mexico, and Thailand",
      "recall_number": "Z-1690-2018",
      "product_description": "ADVIA Centaur¿ HBc IgM",
      "product_quantity": "6",
      "reason_for_recall": "Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.",
      "recall_initiation_date": "20180102",
      "center_classification_date": "20180504",
      "termination_date": "20230210",
      "report_date": "20180516",
      "code_info": "ADVIA Centaur¿ HBc IgM;  test, hepatitis b (b core, be antigen, be antibody, b core igm); SMN 10308978     Lot codes:  ADVIA Centaur aHBcM Lot #  02437164  14331164  20876164  29394164  81764164  93146164  26976167  39604167  44125167  56754167  67635167  77328167  01396170  07110170  21484170  67103170  86117170  92364170  31751171  33726171  41538171    ***Added 9/11/20***  10308978    ***Added 11/2/20***  62572190    ***Added 3/11/21***  39927194"
    }
  ]
}