  "results": [
    {
      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79505",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Canada, Mexico, and Thailand",
      "recall_number": "Z-1689-2018",
      "product_description": "ADVIA Centaur¿ Folate (FOL)",
      "product_quantity": "N/A",
      "reason_for_recall": "Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.",
      "recall_initiation_date": "20180102",
      "center_classification_date": "20180504",
      "termination_date": "20230210",
      "report_date": "20180516",
      "code_info": "ADVIA Centaur¿ Folate; acid, folic, radioimmunoassay; SMN 10310308; 10325366; 10331250     Lot codes:  ADVIA Centaur Folate (REF) Lot #  03460256  09302256  81236256  09294257  30152258  ADVIA Centaur Folate (100 Test) Lot #  76793256  85114256  96368256  05299257  99361257  19045258  25443258  40856260  ADVIA Centaur Folate (500 test) Lot #  73249256  76207256  85113256  00245257  00495257  03779257  05298257  99360257  19042258  19043258  19044258  31130258  36171260"
    }
  ]
}