  "results": [
    {
      "status": "Terminated",
      "city": "Fort Worth",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73883",
      "recalling_firm": "Alcon Research, Ltd.",
      "address_1": "6201 South Fwy",
      "address_2": "N/A",
      "postal_code": "76134-2099",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "United States, Albania, Australia, Brazil, Chile, Costa Rica, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Romania, Slovakia, Spain, Taiwan, Thailand, Turkey, UAE, Ukraine",
      "recall_number": "Z-1688-2016",
      "product_description": "LenSx Laser System.  Indicated for use in patients undergoing cataract surgery and refractive surgery.",
      "product_quantity": "81 units",
      "reason_for_recall": "Alcon is conducting a voluntary recall of all LenSx Lasers with Corneal Flap Capability after receiving reports concerning incomplete corneal flap creation, which manifests as an un-dissected area(s) within the side cut.",
      "recall_initiation_date": "20160414",
      "center_classification_date": "20160517",
      "termination_date": "20171020",
      "report_date": "20160525",
      "code_info": "Serial Numbers: 0511-A013  0313-A345  0713-A455  0415-A019  0912-A258  0313-A359  1213-A613  0412-A142  0412-A137  1114-A943  0212-A112  0313-A357  0813-A469  1113-A605  0314-A697  0315-A981  0913-A530  1012-A286  0515-A037  1112-A291  0515-A042  1011-A065  1012-A284  1013-A555  1211-A086  1012-A287  0912-A245  0614-A783  1113-A596  1014-A916  1011-A062  0615-A044  0913-A506  0313-A353  0615-A050  0615-A056  0912-A247  1114-A938  0712-A202  0714-A824  0113-A320  0212-A117  0614-A782  0614-A784  1112-A294  0712-A201  0514-A749  0715-A071  1111-A067  1214-A948  1013-A542  0314-A687  1213-A614  0215-A976  0312-A135  1012-A285  1013-A548  0215-A978  0412-A139  1113-A586  0312-A129  1011-A050  1213-A612  0314-A685  1112-A292  1012-A264  1015-A126  1111-A077  1013-A547  1114-A931  0112-A099  0112-A101  0413-A372  0413-A380  0512-A157  0612-A200  0715-A073  0811-A034  0815-A097  1013-A554  1113-A587"
    }
  ]
}