  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lisburn",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87804",
      "recalling_firm": "James Leckey Design Ltd",
      "address_1": "19C Ballinderry Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution in the states of CA and WI and the countries of Italy, Norway, Australia, Czechia, Netherlands, Denmark, New Zealand, Japan, United Arab Emirates, Qatar, Croatia, Spain, Poland, Russia, South Africa, Taiwan, Saudi Arabia, Chemitec, Centre Harika, Hong Kong, Lebanon, Belgium, Philippines, Singapore, Malaysia, Indonesia, Slovakia, China, and France.",
      "recall_number": "Z-1687-2021",
      "product_description": "MyWay Walker Chassis - Product Usage: is intended for use by children already using a supportive walking device.",
      "product_quantity": "2,719 walkers",
      "reason_for_recall": "Due to a failure of the welds, the casters detached from the rear legs of the walker.",
      "recall_initiation_date": "20210301",
      "center_classification_date": "20210527",
      "report_date": "20210602",
      "code_info": "Catalog Codes/Item Number:  S1 - 173-1600/PMWS1x;  S2 - 173-2600/PMWS2x ;  S3 - 173-3600/PMWS3x;    Serial Number Range:  220568 to 272111 manufactured between 08/01/2018 to 11/26/2020"
    }
  ]
}