  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79505",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Canada, Mexico, and Thailand",
      "recall_number": "Z-1687-2018",
      "product_description": "ADVIA Centaur¿ Cyclosporine",
      "product_quantity": "N/A",
      "reason_for_recall": "The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere",
      "recall_initiation_date": "20180102",
      "center_classification_date": "20180504",
      "termination_date": "20230210",
      "report_date": "20180516",
      "code_info": "ADVIA Centaur¿ Cyclosporine; Cyclosporine; SMN 10335448    Lot codes:  ADVIA Centaur CSA Lot #  38182025  64432025  98554025  58299030  25547030  30153031  77934031  97170031"
    }
  ]
}