  "results": [
    {
      "status": "Terminated",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65183",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1683-2013",
      "product_description": "Two Lead Arthroscopic Irrigation Set, for use with ARTHROMATIC Plastic Containers, Product Code 2C4030 - Approximate Length 98\" (2.5. m)    Usage:  Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.",
      "product_quantity": "530,488 units",
      "reason_for_recall": "A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.",
      "recall_initiation_date": "20130515",
      "center_classification_date": "20130710",
      "termination_date": "20150616",
      "report_date": "20130717",
      "code_info": "R06K01067  R06K01075  R06K01083  R07B02013  R07C14065  R07C22035  R07D19088  R07E07015  R07E20935  R07E29050  R07F02063  R07F05025  R07I22065  R07J01019  R07J28012  R07L05073  R08A14027  R08A28019  R08B04026  R08B05031  R08C18040  R08C24022  R08D12066  R08D15085  R08F28075  R08G19122  R08G21029  R08G21037  R08H21019  R08H27065  R08I20076  R08J10175  R08K04010  R08L05205  R09A17060  R09A28109  R09B16086  R09C16100  R09C30119  R09D14012  R09E19026  R09E25122  R09F10148  R09F16087  R09G14072  R09G27025  R09H27080  R09I28094  R09J13103  R09K11154  R09K30014  R09L09032  R09L16227  R10A08264  R10A19188  R10A21630  R10B13122  R10B16158  R10C10118  R10C13096  R10C24044  R10C27179  R10D05041  R10D26047  R10E11087  R10F05152  R10F19054  R10G27014  R10H17211  R10H31212  R10I01106  R10I07129  R10I27036  R10J19024  R10J19149  R10J23067  R10K02226  R10K13066  R10K25029  R10L04071  R10L23048  R11A11010  R11A15078  R11A31224  R11B07172  R11C01223  R11C15157  R11C22294  R11D04167  R11D16062  R11E12159  R11E21077  R11F24103  R11G18020  R11G18103  R11G19077  R11J24189  R11K22140  R11L22114  R12A04113  R12A11134  R12A23113  R12A31124  R12B15117  R12B27070  R12C08144  R12E17116  R12E18056  R12F07057  R12F09087  R12F11067"
    }
  ]
}