  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom.  This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.",
      "address_2": "",
      "product_quantity": "10 units",
      "code_info": "Serial #'s: 11005, 11006, 11007, 11008, 11009, 11010, 11011, 11012, 11013, 11014 & 24061.",
      "center_classification_date": "20150528",
      "distribution_pattern": "Worldwide Distribution -- US, including the states of NY, OH, & TX; and, the countries of Finland, Germany, Japan, Netherlands, South Korea, Spain, & Switzerland.",
      "state": "OH",
      "product_description": "Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems.    The device is a diagnostic imaging system for fixed installations that combines Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI).",
      "report_date": "20150603",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-1682-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71228",
      "termination_date": "20180626",
      "more_code_info": "",
      "recall_initiation_date": "20150311",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}