  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65183",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1682-2013",
      "product_description": "Four Lead TUR irrigation Set, for use with UROMATIC Plastic Containers,  Product Code 2C4013 - Approximate Length 85\" (2.2. m)    Usage:  Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.",
      "product_quantity": "356,728 units",
      "reason_for_recall": "A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.",
      "recall_initiation_date": "20130515",
      "center_classification_date": "20130710",
      "termination_date": "20150616",
      "report_date": "20130717",
      "code_info": "UR175554  UR176743  UR182030  UR188144  UR192047  UR192096  UR192260  UR194951  UR202044  UR204032  UR211698  UR214312  UR216069  UR222471  UR225334  UR229294  UR233569  UR233700  UR238915  UR243022  UR246009  UR249383  UR255802  UR260273  UR262980  UR263004  UR265918  UR283580  UR287284  UR288746  UR298216  UR304246  UR307249  UR314328  UR318063  UR321000  UR325464  UR334151  UR339317  UR344234  UR352484  UR352567  UR360255  UR363432  UR366161  UR375857  UR383794  UR385534  UR07J19029  UR07K29059  UR07L21039  UR08B21053  UR08C25151  UR08E15157  UR08G22167  UR08I12065  UR08K13086  UR08L18158  UR09A21211  UR09D02157  UR09E14084  UR09G17125  UR09H26140  UR09J30040  UR09J30065  UR09J30107  UR09L01146  UR10C25033  UR10D09142  UR10E21037  UR10G28129  UR10I28034  UR10L03024  UR10L22024  UR11A14022  UR11B24086  UR11B25125  UR11C25032  UR11F01028  UR11G08088  UR11H08011  UR11H09019  UR11I08043  UR11I23026  UR12A25043  UR12E18502  UR12H03011  UR12H17045  UR12I05030  UR12I06053  UR12I21052  UR12J18040  UR12J26019  UR12L12023  UR13B08053  UR13B15041  UR13C08036"
    }
  ]
}