{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79505",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and Canada, Mexico, and Thailand",
      "recall_number": "Z-1680-2018",
      "product_description": "IMMULITE ¿ /IMMULITE 1000 ¿ Gastrin",
      "product_quantity": "N/A",
      "reason_for_recall": "Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.",
      "recall_initiation_date": "20180102",
      "center_classification_date": "20180504",
      "termination_date": "20230210",
      "report_date": "20180516",
      "code_info": "Gastrin, Gastrin (IMMULITE/IMMULITE 1000, LKGA1), (IMMULITE 2000/IMMULITE 2000 XPi, LKGA2)    Lot codes:  Gastrin Lot #  0345  0346  0347"
    }
  ]
}