  "results": [
    {
      "status": "Terminated",
      "city": "Largo",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65320",
      "recalling_firm": "Linvatec Corp. dba ConMed Linvatec",
      "address_1": "11311 Concept Blvd",
      "address_2": "N/A",
      "postal_code": "33773-4908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.",
      "recall_number": "Z-1680-2013",
      "product_description": "***REF C7120***APEX Arthroscopy Tubing Set.    Intended for use for joint irrigation and dissention during arthroscopic procedures.",
      "product_quantity": "4,691 Units total of each",
      "reason_for_recall": "Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.",
      "recall_initiation_date": "20130530",
      "center_classification_date": "20130710",
      "termination_date": "20130904",
      "report_date": "20130717",
      "code_info": "C7120 Lot's #: 1301184, 1301214, 1301254, 1301284, 1301294, 1302014, 1302044, 1302054, 1302064, 1302074, 13020124, 1302134, 1302144, 1302194, 1302204, 1302214, 1302224, 1302254, 1302264, 1303014, 1303054, 1303064, 1303074, 1303114, 1303124, 1303204, 1303214, 1303264, 1303274, 1303284, 1304014, 1304034, 1304054, 1304154, 1304294, 1304304, and 1305014."
    }
  ]
}