{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85161",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia, Wyoming",
      "recall_number": "Z-1679-2020",
      "product_description": "ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13",
      "product_quantity": "1012",
      "reason_for_recall": "Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.",
      "recall_initiation_date": "20200218",
      "center_classification_date": "20200407",
      "termination_date": "20210209",
      "report_date": "20200415",
      "code_info": "Serial Numbers  3905,  70-4077,  4082,  70-4283,  70-4368,  5177,  3922,  5120,  5170,  5179,  5205,  5226,  5245,  5282,  5300,  5343,  5385,  5391,  5506,  5665,  4078,  70-4143,  70-4235,  70-4296,  70-4344,  5154,  5281,  5417,  3755,  3761,  3802,  3801,  3901,  3981,  4072,  70-4174,  70-4354,  70-4378,  5087,  5088,  5096,  5095,  5098,  5118,  5148,  5396,  5157,  5821"
    }
  ]
}